. 2023 Sep;90(3):1010-1024.
doi: 10.1002/mrm.29689. Epub 2023 May 28.

A multiparametric alternative to short inversion-time inversion recovery for imaging inflammation: T2water and fat fraction measurement using chemical shift-encoded turbo spin-echo MRI

Affiliations

A multiparametric alternative to short inversion-time inversion recovery for imaging inflammation: T2water and fat fraction measurement using chemical shift-encoded turbo spin-echo MRI

Ruaridh M Gollifer et al. Magn Reson Med. 2023 Sep.

Abstract

Purpose: Short-inversion-time inversion-recovery MRI is used widely for imaging bone and soft-tissue inflammation in rheumatic inflammatory diseases, but there is no widely available quantitative equivalent of this sequence. This limits our ability to objectively assess inflammation and distinguish it from other processes. To address this, we investigate the use of the widely available Dixon turbo spin echo (TSE Dixon) sequence as a practical approach to simultaneous water-specific T2 (T2water ) and fat fraction (FF) measurement.

Methods: We use a series of TSE Dixon acquisitions with varying effective TEs (TEeff ) to quantify T2water and FF. The validity of this approach is assessed in a series of phantom and in vivo experiments, with reference values provided by Carr-Purcell-Meiboom-Gill acquisitions, MRS, and phantoms. The effect of inflammation on parameter values is evaluated in patients with spondyloarthritis.

Results: The T2water estimates obtained from TSE Dixon were accurate compared with the reference values from Carr-Purcell-Meiboom-Gill and spectroscopy in both fat-free environments and in the presence of fat. FF measurements with T2water correction from TSE Dixon were accurate from 0% to 60% FF and were not confounded by T2water variations. In vivo imaging produced good quality images that were artifact free, produced plausible T2 values, separating and quantifying the effect of inflammation on T2water and FF.

Conclusion: The T2water and FF measurements based on TSE Dixon with effective TE increments are accurate over a range of T2 and FF values and could provide a widely available quantitative alternative to the short-inversion-time inversion-recovery sequence for imaging inflamed tissue.

Keywords: inflammation; magnetic resonance imaging; radiology.

PubMed Disclaimer

References

REFERENCES

    1. Walsh NC, Crotti TN, Goldring SR, et al. Rheumatic diseases: the effects of inflammation on bone. Immunol Rev. 2005;208:228-251. doi:10.1111/J.0105-2896.2005.00338.X
    1. Sieper J, Rudwaleit M, Baraliakos X, et al. The assessment of Spondyloarthritis International Society (ASAS) Handbook: a guide to assess spondyloarthritis. Ann Rheum Dis. 2009;68:ii1-ii44. doi:10.1136/ard.2008.104018
    1. Bray TJP, Jones A, Bennett AN, et al. Recommendations for acquisition and interpretation of MRI of the spine and sacroiliac joints in the diagnosis of axial spondyloarthritis in the UK. Rheumatology. 2019;58:1831-1838. doi:10.1093/rheumatology/kez173
    1. Day J, Patel S, Limaye V. The role of magnetic resonance imaging techniques in evaluation and management of the idiopathic inflammatory myopathies. Semin Arthritis Rheum. 2017;46:642-649. doi:10.1016/J.SEMARTHRIT.2016.11.001
    1. Miller E, Uleryk E, Doria AS. Evidence-based outcomes of studies addressing diagnostic accuracy of MRI of juvenile idiopathic arthritis. Am J Roentgenol. 2012;192:1209-1218. doi:10.2214/AJR.08.2304

Publication types

LinkOut - more resources