Accelerated Hemocompatibility Testing of Rotary Blood Pumps
- PMID: 37256782
- DOI: 10.1097/MAT.0000000000001995
Accelerated Hemocompatibility Testing of Rotary Blood Pumps
Abstract
Ex vivo hemocompatibility testing is a vital element of preclinical assessment for blood-contacting medical devices. Current approaches are resource intensive; thus, we investigated the feasibility of accelerating hemocompatibility testing by standardizing the number of pump exposures in loops of various sizes. Three identical blood loops were constructed, each with a custom-molded reservoir able to facilitate large-volume expansion. Using the HVAD rotary blood pump operating at 5 L·min -1 and 100 mmHg, three test volumes (80, 160, and 320 ml) were circulated for 4000 pump exposures. Blood sampling was performed at individualized intervals every one-sixth of total duration for the assessment of hemolysis and von Willebrand Factor (vWF) degradation. While steady increases in hemolysis (~24 mg·dl -1 ) were identified in all tests at completion, loop volume was not a primary discriminator. The normalized index of hemolysis did not vary significantly between loops (4.2-4.9 mg·100 L -1 ). vWF degradation progressively occurred with duration of testing to a similar extent under all conditions. These data support an accelerated approach to preclinical assessment of ex vivo blood damage. Adopting this approach enables: enhanced efficiency for rapid prototyping; reduced ex vivo blood aging, and; greater utility of blood, which is presently limited if 450 ml loops are desired.
Copyright © ASAIO 2023.
Conflict of interest statement
Disclosure: The authors have no conflicts of interest to report.
References
-
- ASTM Standard F1841, 19e1. Standard Practice for Assessment of Hemolysis in Continuous Flow Blood Pumps. ASTM International, West Conshohocken, PA, 2019. doi:10.1520/F1841-19E01.
-
- Kirklin JK, Pagani FD, Kormos RL, et al.: Eighth annual INTERMACS report: Special focus on framing the impact of adverse events. J Heart Lung Transplant 36: 1080–1086, 2017.
-
- Kuehn BM: FDA: Stop using Medtronic’s Heartware ventricular assist device. JAMA 326: 215–215, 2021.
-
- ASTM Standard F1830, 19. Standard practice for collection and preparation of blood for dynamic in vitro evaluation of hemolysis in blood pumps. ASTM International, West Conshohocken, PA, 2019. doi: 10.1520/F1830-19.
-
- McNamee AP, Kuck L, Simmonds MJ: Bovine erythrocytes are poor surrogates for human when exposed to sublethal shear stress. Int J Artif Organs: 45: 580–587, 2022.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous