. 2024 Feb;59(2):661-672.
doi: 10.1002/jmri.28834. Epub 2023 Jun 1.

Quantification of Cerebral Glucose Concentrations via Detection of the H1-α-Glucose Peak in 1 H MRS at 7 T

Affiliations

Quantification of Cerebral Glucose Concentrations via Detection of the H1-α-Glucose Peak in 1 H MRS at 7 T

Hideto Kuribayashi et al. J Magn Reson Imaging. 2024 Feb.

Abstract

Background: Sensitive detection and quantification of cerebral glucose is desired.

Purpose: To quantify cerebral glucose by detecting the H1-α-glucose peak at 5.23 ppm in 1 H magnetic resonance spectroscopy at 7 T.

Study type: Prospective.

Subjects: Twenty-eight non-fasted healthy subjects (aged 20-28 years).

Field strength/sequence: Short echo time stimulated echo acquisition mode (short-TE STEAM) and semi-localized by adiabatic selective refocusing (semi-LASER) at 7 T.

Assessment: Single voxel spectra were obtained from the posterior cingulate cortex (27-mL) using a 32-channel head coil. The H1-α-glucose peak in the spectrum with retrospective removal of the residual water peak was fitted using LCModel with a glucose basis set of only the H1-α-glucose peak. Conventional spectral analysis was performed with a glucose basis set of a full spectral pattern of glucose, also. Fitting precision was evaluated with Cramér-Rao lower bounds (CRLBs). The repeatability of glucose quantification via the semi-LASER sequence was tested.

Statistical tests: Paired or Welch's t-test were used for normally distributed values. A P value of <0.05 was considered significant. The repeatability of measures was analyzed using coefficient of variation (CV).

Results: Removal of the residual water peak improved the flatness and stability of baselines around the H1-α-glucose peak and reduced CRLBs for fitting the H1-α-glucose peak. The semi-LASER sequence was superior to the short-TE STEAM in the higher signal-to-noise ratio of the H1-α-glucose peak (mean ± SD 7.9 ± 2.5, P < 0.001). The conventional analysis overfitted the H1-α-glucose peak. The individual CVs of glucose quantification by detecting the H1-α-glucose peak were smaller than the corresponding CRLBs.

Data conclusion: Cerebral glucose concentration is quantitated to be 1.07 mM by detecting the H1-α-glucose peak in the semi-LASER spectra. Despite requiring long scan times, detecting the H1-α-glucose peak allows true glucose quantification free from the influence of overlapping taurine and macromolecule signals.

Evidence level: 2 TECHNICAL EFFICACY STAGE: 1.

Keywords: 7 T; H1-α-glucose; LCModel; baseline correction; brain glucose concentration; magnetic resonance spectroscopy.

PubMed Disclaimer

Comment in

References

    1. Gruetter R, Novotny EJ, Boulware SD, et al. Direct measurement of brain glucose concentrations in humans by 13C NMR spectroscopy. Proc Natl Acad Sci U S A 1992;89:1109-1112.
    1. Gruetter R, Garwood M, Ugurbil K, Seaquist ER. Observation of resolved glucose signals in 1H NMR spectra of the human brain at 4T. Magn Reson Med 1996;36:1-6.
    1. Gruetter R, Ugurbil K, Seaquist ER. Steady-state cerebral glucose concentrations and transport in the human brain. J Neurochem 1998;70:397-408.
    1. Duarte JMN, Gruetter R. Cerebral glucose transport and homeostasis. In: Choi IY, Gruetter R, editors. Neural metabolism in vivo. New York: Springer; 2012. p 655-673.
    1. Joers JM, Deelchand DK, Kumar A, et al. Measurement of hypothalamic glucose under euglycemia and hyperglycemia by MRI at 3T. J Magn Reson Imaging 2017;45:681-691.

Publication types

LinkOut - more resources