Performance of a multisensor implantable defibrillator algorithm for heart failure monitoring related to co-morbidities
- PMID: 37278122
- PMCID: PMC10375157
- DOI: 10.1002/ehf2.14416
Performance of a multisensor implantable defibrillator algorithm for heart failure monitoring related to co-morbidities
Abstract
Aims: The HeartLogic algorithm combines multiple implantable defibrillator (ICD) sensor data and has proved to be a sensitive and timely predictor of impending heart failure (HF) decompensation in cardiac resynchronization therapy (CRT-D) patients. We evaluated the performance of this algorithm in non-CRT ICD patients and in the presence of co-morbidities.
Methods and results: The HeartLogic feature was activated in 568 ICD patients (410 with CRT-D) from 26 centres. The median follow-up was 26 months [25th-75th percentile: 16-37]. During follow-up, 97 hospitalizations were reported (53 cardiovascular) and 55 patients died. We recorded 1200 HeartLogic alerts in 370 patients. Overall, the time IN the alert state was 13% of the total observation period. The rate of cardiovascular hospitalizations or death was 0.48/patient-year (95% CI: 0.37-0.60) with the HeartLogic IN the alert state and 0.04/patient-year (95% CI: 0.03-0.05) OUT of the alert state, with an incidence rate ratio of 13.35 (95% CI: 8.83-20.51, P < 0.001). Among patient characteristics, atrial fibrillation (AF) on implantation (HR: 1.62, 95% CI: 1.27-2.07, P < 0.001) and chronic kidney disease (CKD) (HR: 1.53, 95% CI: 1.21-1.93, P < 0.001) independently predicted alerts. HeartLogic alerts were not associated with CRT-D versus ICD implantation (HR: 1.03, 95% CI: 0.82-1.30, P = 0.775). Comparisons of the clinical event rates in the IN alert state with those in the OUT of alert state yielded incidence rate ratios ranging from 9.72 to 14.54 (all P < 0.001) in all groups of patients stratified by: CRT-D/ICD, AF/non-AF, and CKD/non-CKD. After multivariate correction, the occurrence of alerts was associated with cardiovascular hospitalization or death (HR: 1.92, 95% CI: 1.05-3.51, P = 0.036).
Conclusions: The burden of HeartLogic alerts was similar between CRT-D and ICD patients, while patients with AF and CKD seemed more exposed to alerts. Nonetheless, the ability of the HeartLogic algorithm to identify periods of significantly increased risk of clinical events was confirmed, regardless of the type of device and the presence of AF or CKD.
Trial registration: ClinicalTrials.gov NCT02275637.
Keywords: Atrial fibrillation; CRT; Chronic kidney disease; Heart failure; ICD; Risk stratification.
© 2023 The Authors. ESC Heart Failure published by John Wiley & Sons Ltd on behalf of European Society of Cardiology.
Conflict of interest statement
M. Campari and S. Valsecchi are employees of Boston Scientific, Inc. No other conflicts of interest exist.
Figures
References
-
- McDonagh TA, Metra M, Adamo M, Gardner RS, Baumbach A, Böhm M, Burri H, Butler J, Čelutkienė J, Chioncel O, Cleland JGF, Coats AJS, Crespo‐Leiro MG, Farmakis D, Gilard M, Heymans S, Hoes AW, Jaarsma T, Jankowska EA, Lainscak M, Lam CSP, Lyon AR, McMurray JJV, Mebazaa A, Mindham R, Muneretto C, Francesco Piepoli M, Price S, Rosano GMC, Ruschitzka F, Kathrine Skibelund A, ESC Scientific Document Group . ESC scientific document group. 2021 ESC guidelines for the diagnosis and treatment of acute and chronic heart failure. Eur Heart J. 2021; 42: 3599–3726. - PubMed
-
- Goldenberg I, Huang DT, Nielsen JC. The role of implantable cardioverter‐defibrillators and sudden cardiac death prevention: indications, device selection, and outcome. Eur Heart J. 2020; 41: 2003–2011. - PubMed
-
- Guaricci AI, Masci PG, Muscogiuri G, Guglielmo M, Baggiano A, Fusini L, Lorenzoni V, Martini C, Andreini D, Pavon AG, Aquaro GD, Barison A, Todiere G, Rabbat MG, Tat E, Raineri C, Valentini A, Varga‐Szemes A, Schoepf UJ, de Cecco CN, Bogaert J, Dobrovie M, Symons R, Focardi M, Gismondi A, Lozano‐Torres J, Rodriguez‐Palomares JF, Lanzillo C, di Roma M, Moro C, di Giovine G, Margonato D, de Lazzari M, Perazzolo Marra M, Nese A, Casavecchia G, Gravina M, Marzo F, Carigi S, Pica S, Lombardi M, Censi S, Squeri A, Palumbo A, Gaibazzi N, Camastra G, Sbarbati S, Pedrotti P, Masi A, Carrabba N, Pradella S, Timpani M, Cicala G, Presicci C, Puglisi S, Sverzellati N, Santobuono VE, Pepi M, Schwitter J, Pontone G. CarDiac magnEtic Resonance for prophylactic Implantable‐cardioVerter defibrillAtor ThErapy in Non‐Ischaemic dilated CardioMyopathy: an international Registry. Europace. 2021; 23: 1072–1083. - PubMed
-
- Conraads VM, Tavazzi L, Santini M, Oliva F, Gerritse B, Yu CM, Cowie MR. Sensitivity and positive predictive value of implantable intrathoracic impedance monitoring as a predictor of heart failure hospitalizations: the SENSE‐HF trial. Eur Heart J. 2011; 32: 2266–2273. - PubMed
-
- van Veldhuisen DJ, Braunschweig F, Conraads V, Ford I, Cowie MR, Jondeau G, Kautzner J, Muñoz Aguilera R, Lunati M, Yu CM, Gerritse B, Borggrefe M, for the DOT‐HF Investigators . DOT‐HF investigators. Intrathoracic impedance monitoring, audible patient alerts, and outcome in patients with heart failure. Circulation. 2011; 124: 1719–1726. - PubMed