Extensive misalignment of plate-haptic rotationally asymmetric multifocal toric intraocular lens

Affiliations

Extensive misalignment of plate-haptic rotationally asymmetric multifocal toric intraocular lens

Tetsuro Oshika et al. Jpn J Ophthalmol. 2023 Sep.

Abstract

Purpose: Plate-haptic rotationally asymmetric multifocal toric intraocular lenses (IOL) (Lentis Comfort Toric) occasionally rotate extensively after surgery. We conducted the current study to investigate the incidence of extensive misalignment of this IOL and its association with clinical parameters.

Study design: Retrospective case series.

Subjects and methods: Data were collected from patients who had undergone phacoemulsification and implantation of a plate-haptic multifocal toric IOL.

Results: Among 332 eyes, extensive misalignment of toric IOLs ≥ 45º occurred in 3.3% (11 eyes). The amount of misalignment in eyes with extensive misalignment was 81.6 ± 22.9º, while in those without extensive misalignment, it was 3.0 ± 2.7º. The eyes with extensive misalignment showed significantly greater axial length (p < 0.001), larger corneal diameter (p = 0.034), and flatter corneas (p = 0.044) than those without extensive misalignment. Repositioning surgery to correct toric IOL misorientation was conducted in 9 eyes between 7 and 28 days after cataract surgery. In 2 eyes, repositioning surgery was carried out twice.

Conclusions: In majority of cases plate-haptic multifocal toric IOLs showed satisfactory rotational stability, but extensive misalignment ≥ 45º occurred in 3.3% of cases.

Keywords: Axis misalignment; Multifocal toric intraocular lens; Plate-haptic lens; Rotational stability.

PubMed Disclaimer

References

    1. Lee BS, Chang DF. Comparison of the rotational stability of two toric intraocular lenses in 1273 consecutive eyes. Ophthalmology. 2018;125:1325–31. - DOI - PubMed
    1. Kramer BA, Hardten DR, Berdahl JP. Rotation characteristics of three toric monofocal intraocular lenses. Clin Ophthalmol. 2020;14:4379–84. - DOI - PubMed
    1. Oshika T, Inamura M, Inoue Y, Ohashi T, Sugita T, Fujita Y, et al. Incidence and outcomes of repositioning surgery to correct misalignment of toric intraocular lenses. Ophthalmology. 2018;125:31–5. - DOI - PubMed
    1. Lee BS, Onishi AC, Chang DF. Comparison of rotational stability and repositioning rates of 2 presbyopia-correcting and 2 monofocal toric intraocular lenses. J Cataract Refract Surg. 2021;47:622–6. - DOI - PubMed
    1. Takaku R, Nakano S, Iida M, Oshika T. Influence of frosted haptics on rotational stability of toric intraocular lenses. Sci Rep. 2021;11:15099. - DOI - PubMed

LinkOut - more resources