. 2023 Aug:144:105932.
doi: 10.1016/j.jmbbm.2023.105932. Epub 2023 May 26.

Metal debris release is commonly seen from explanted total ankle replacements

Affiliations
Free article

Metal debris release is commonly seen from explanted total ankle replacements

Shona Haston et al. J Mech Behav Biomed Mater. 2023 Aug.
Free article

Abstract

This study aimed to characterise the damage mechanisms present on the metal components used in various contemporary total ankle replacements. Twenty-seven explanted total ankle replacements comprising 8 different designs (3 fixed bearing and 5 mobile bearing) were analysed using various explant analysis techniques. Pitting and scratching were the most commonly observed wear features. Microscopic analysis revealed metallic pitting on 52% of tibial components and 95% of talar components. Pitting was identified on more cobalt-chromium than titanium alloy tibial components (63% versus 0%). Non-contact profilometry confirmed the presence of pitting, with significant (p < 0.05) differences in the measured average surface roughness values of pitted and unpitted areas for tibial and talar components. There was macroscopically visible sliding plane scratching, indicating the presence of hard third body particles, on 78% of talar components. Changes to the non-articulating surfaces coatings in terms of coating loss and/or changes in reflectivity was identified visually on 80% of metal components. Scanning electron microscopy with energy dispersive X-ray spectroscopy identified metallic embedded debris in 19% of polyethylene inserts. This explant study demonstrates the release of metal debris from both the metallic tibial and talar component articulating surfaces and non-articulating surface coatings of various contemporary total ankle replacements. Metal particulate debris release from total ankle replacements may be more common than previously recognised. Metal debris should be considered in further study into the aetiology of failed total ankle arthroplasty.

Keywords: Explant analysis; Metal release; Total ankle replacement; Wear.

PubMed Disclaimer

Conflict of interest statement

Declaration of competing interest The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: David Langton is an expert witness for plaintiffs in ongoing metal-on-metal hip litigation. David Townshend reports institutional support and personal consultancy fees, both for Stryker.