Reliability of Respiratory System Compliance Calculation During Assisted Mechanical Ventilation: A Retrospective Study
- PMID: 37326475
- DOI: 10.1097/CCM.0000000000005964
Reliability of Respiratory System Compliance Calculation During Assisted Mechanical Ventilation: A Retrospective Study
Abstract
Objectives: To compare respiratory system compliance (C rs ) calculation during controlled mechanical ventilation (MV) and, subsequently, during assisted MV.
Design: This is a single-center, retrospective, observational study.
Setting: This study was conducted on patients admitted to Neuro-ICU of Niguarda Hospital (tertiary referral hospital).
Patients: We analyzed every patient greater than or equal to 18 years old having a C rs measurement in controlled and in assisted MV within 60 minutes. Plateau pressure (P plat ) was considered reliable if it was deemed visually stable for at least 2 seconds.
Interventions: Inspiratory pause was incorporated to detect P plat in controlled and assisted MV. Calculation of C rs and driving pressure were achieved.
Measurements and main results: A total of 101 patients were studied. An acceptable agreement was found (Bland-Altman plot bias -3.9, level of agreement upper 21.6, lower -29.6). C rs in assisted MV was 64.1 (52.6-79.3) and in controlled MV it was 61.2 (50-71.2) mL/cm H 2o ( p = 0.006). No statistical difference was found in C rs (assisted vs controlled MV) when peak pressure was lower than P plat nor when peak pressure was higher than P plat .
Conclusions: A P plat visually stable for at least 2 seconds leads to reliable C rs calculation during assisted MV.
Copyright © 2023 by the Society of Critical Care Medicine and Wolters Kluwer Health, Inc. All Rights Reserved.
Conflict of interest statement
Dr. Fumagalli disclosed work for hire. The remaining authors have disclosed that they do not have any potential conflicts of interest.
Comment in
-
Is Respiratory System Compliance Calculation During Assisted Mechanical Ventilation Reliable?Crit Care Med. 2024 Jan 1;52(1):e27-e28. doi: 10.1097/CCM.0000000000006036. Epub 2023 Dec 14. Crit Care Med. 2024. PMID: 38095528 No abstract available.
References
-
- Russotto V, Bellani G, Foti G: Respiratory mechanics in patients with acute respiratory distress syndrome. Ann Transl Med 2018; 6:382
-
- Grasselli G, Brioni M, Zanella A: Monitoring respiratory mechanics during assisted ventilation. Curr Opin Crit Care 2020; 26:11–17
-
- Bellani G, Grassi A, Sosio S, et al.: Plateau and driving pressure in the presence of spontaneous breathing. Intensive Care Med 2019; 45:97–98
-
- Bianchi I, Grassi A, Pham T, et al.: Reliability of plateau pressure during patient-triggered assisted ventilation. Analysis of a multicentre database [published correction appears in J Crit Care. 2022 Apr 1;:154030]. J Crit Care 2022; 68:96–103
-
- Rezoagli E, Bastia L, Grassi A, et al.: Paradoxical effect of chest wall compression on respiratory system compliance: A multicenter case series of patients with ARDS, with multimodal assessment. Chest 2021; 160:1335–1339
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous