Inter- and intrarater reliability of short-term measurement of heart rate variability on rest in chronic obstructive pulmonary disease (COPD)

Affiliations

Inter- and intrarater reliability of short-term measurement of heart rate variability on rest in chronic obstructive pulmonary disease (COPD)

Aldair Darlan Santos-de-Araújo et al. Heart Lung. 2023 Nov-Dec.

Abstract

Background: Individuals with COPD have an imbalance of cardiac autonomic control. In this context, HRV is considered an important tool for assessing cardiac sympathetic and parasympathetic balance, however it is a dependent evaluator measure and subject to methodological biases that may compromise the interpretation of results.

Objective: This study examines the inter- and intrarater reliability of HRV parameters derived from short-term recordings in individuals with COPD.

Methodology: Fifty-one individuals of both genders with COPD clinical diagnosis confirmed by the pulmonary function test and aged ≥50 years were included. The RR interval (RRi) were recorded during a 10 min period on supine position using a portable heart rate monitor (Polar® H10 model). The data were transferred into Kubios® HRV Standard analysis software and analyzed within the stable sessions containing 256 sequential RRi.

Results: The intraclass correlation coefficient (ICC) ranged from 0.942 to 1.000 according to the intrarater analysis by Researcher 01 and 0.915 to 0.998 to the intrarater by Researcher 02. The interrater ICC ranged from 0.921 to 0.998. The coefficient of variation was up to 8.28 for Researcher 01 intrarater analysis, 9.06 for Researcher 02 intrarater analysis and 13.07 for interrater analysis.

Conclusion: The measurement of HRV using a portable heart rate device in individuals with COPD present acceptable values of intra- and interrater reliability, supporting the use of HRV in the clinical and scientific scenario. Furthermore, it is important that the data analysis be performed by the same experienced evaluator.

Keywords: Chronic Obstructive Pulmonary Disease; Heart Rate Variability; Kubios Software; Reliability.