Biomechanical simulation of forces and moments of initial orthodontic tooth movement in dependence on the used archwire system by ROSS (Robot Orthodontic Measurement & Simulation System)
- PMID: 37348170
- DOI: 10.1016/j.jmbbm.2023.105960
Biomechanical simulation of forces and moments of initial orthodontic tooth movement in dependence on the used archwire system by ROSS (Robot Orthodontic Measurement & Simulation System)
Abstract
Objectives: Aim of this study was to determine the forces and moments during simulated initial orthodontic tooth movements using a novel biomechanical test setup.
Methods: The test setup consisted of an industrial precision robot with a force-torque sensor, a maxillary model and a control computer and software. Forces and moments acting on the corresponding experimental tooth during the motion simulations were dynamically measured for two 0.016" NiTi round archwires (Sentalloy Light/Sentalloy Medium). Intrusive (#1), rotational (#2) and angular (#3) tooth movements were simulated by a control program based on the principle of force control and executed by the robot. The results were statistically analysed using K-S-test and Mann-Whitney U test with a significance level of α = 5%.
Results: Sentalloy Medium archwires generated higher forces and moments than the Sentalloy Light archwires in all simulations. In simulation #1 the mean initial forces/moments reached 1.442 N/6.781 Nmm for the Light archwires and 1.637 N/9.609 Nmm for the Medium archwires. In movement #2 Light archwires generated mean initial forces/moments of 0.302 N/-8.271 Nmm whereas Medium archwires generated 0.432 N/-9.653 Nmm. Simulation #3 showed mean initial forces/moments of -0.122 N/8.477 Nmm from the Light archwires compared to -0.300 N/11.486 Nmm for the Medium archwires.
Significance: The measured forces and moments were suitable for initial orthodontic tooth movement in simulations #2 and #3, however inadequate in simulation #1. Reduced archwire dimensions (<0.016″) should be selected for initial leveling of vertical malocclusions.
Keywords: Biomechanics; Force control; Leveling archwire; Orthodontic simulation; Orthodontic tooth movement; Robotics.
Copyright © 2023 The Authors. Published by Elsevier Ltd.. All rights reserved.
Conflict of interest statement
Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Similar articles
-
Investigation of Forces and Moments during Orthodontic Tooth Intrusion Using Robot Orthodontic Measurement and Simulation System (ROSS).Bioengineering (Basel). 2023 Nov 25;10(12):1356. doi: 10.3390/bioengineering10121356. Bioengineering (Basel). 2023. PMID: 38135947 Free PMC article.
-
Measurement of forces and moments around the maxillary arch for treatment of a simulated lingual incisor and high canine malocclusion using straight and mushroom archwires in fixed lingual appliances.Eur J Orthod. 2017 Nov 30;39(6):665-672. doi: 10.1093/ejo/cjx028. Eur J Orthod. 2017. PMID: 28430887
-
A comparative assessment of the forces and moments generated at the maxillary incisors between conventional and self-ligating brackets using a reverse curve of Spee NiTi archwire.Aust Orthod J. 2010 Nov;26(2):127-33. Aust Orthod J. 2010. PMID: 21175021
-
Recent Advances in Orthodontic Archwires: A Review.Cureus. 2023 Oct 25;15(10):e47633. doi: 10.7759/cureus.47633. eCollection 2023 Oct. Cureus. 2023. PMID: 38022289 Free PMC article. Review.
-
Multi-Force Bio-Active™ Archwires and Various Contemporary NiTi Multi-Force Archwires: Properties and Characteristics-A Review.Materials (Basel). 2024 May 28;17(11):2603. doi: 10.3390/ma17112603. Materials (Basel). 2024. PMID: 38893867 Free PMC article. Review.
Cited by
-
Seeking orderness out of the orderless movements: an up-to-date review of the biomechanics in clear aligners.Prog Orthod. 2024 Nov 18;25(1):44. doi: 10.1186/s40510-024-00543-1. Prog Orthod. 2024. PMID: 39551881 Free PMC article. Review.
-
Inclination Changes in Incisors During Orthodontic Treatment with Passive Self-Ligating Brackets.J Clin Med. 2025 May 12;14(10):3370. doi: 10.3390/jcm14103370. J Clin Med. 2025. PMID: 40429365 Free PMC article.
-
Application of Robotics in Orthodontics: A Systematic Review.Cureus. 2024 Apr 18;16(4):e58555. doi: 10.7759/cureus.58555. eCollection 2024 Apr. Cureus. 2024. PMID: 38765377 Free PMC article. Review.
-
Biomechanical Simulation of Orthodontic En-Bloc Retraction Comparing Compound Technique and Sliding Mechanics Using a HOSEA Robotic Device.Bioengineering (Basel). 2024 Feb 2;11(2):153. doi: 10.3390/bioengineering11020153. Bioengineering (Basel). 2024. PMID: 38391639 Free PMC article.
-
Investigation of Forces and Moments during Orthodontic Tooth Intrusion Using Robot Orthodontic Measurement and Simulation System (ROSS).Bioengineering (Basel). 2023 Nov 25;10(12):1356. doi: 10.3390/bioengineering10121356. Bioengineering (Basel). 2023. PMID: 38135947 Free PMC article.
MeSH terms
Substances