Biomechanical simulation of forces and moments of initial orthodontic tooth movement in dependence on the used archwire system by ROSS (Robot Orthodontic Measurement & Simulation System)
- PMID: 37348170
- DOI: 10.1016/j.jmbbm.2023.105960
Biomechanical simulation of forces and moments of initial orthodontic tooth movement in dependence on the used archwire system by ROSS (Robot Orthodontic Measurement & Simulation System)
Abstract
Objectives: Aim of this study was to determine the forces and moments during simulated initial orthodontic tooth movements using a novel biomechanical test setup.
Methods: The test setup consisted of an industrial precision robot with a force-torque sensor, a maxillary model and a control computer and software. Forces and moments acting on the corresponding experimental tooth during the motion simulations were dynamically measured for two 0.016" NiTi round archwires (Sentalloy Light/Sentalloy Medium). Intrusive (#1), rotational (#2) and angular (#3) tooth movements were simulated by a control program based on the principle of force control and executed by the robot. The results were statistically analysed using K-S-test and Mann-Whitney U test with a significance level of α = 5%.
Results: Sentalloy Medium archwires generated higher forces and moments than the Sentalloy Light archwires in all simulations. In simulation #1 the mean initial forces/moments reached 1.442 N/6.781 Nmm for the Light archwires and 1.637 N/9.609 Nmm for the Medium archwires. In movement #2 Light archwires generated mean initial forces/moments of 0.302 N/-8.271 Nmm whereas Medium archwires generated 0.432 N/-9.653 Nmm. Simulation #3 showed mean initial forces/moments of -0.122 N/8.477 Nmm from the Light archwires compared to -0.300 N/11.486 Nmm for the Medium archwires.
Significance: The measured forces and moments were suitable for initial orthodontic tooth movement in simulations #2 and #3, however inadequate in simulation #1. Reduced archwire dimensions (<0.016″) should be selected for initial leveling of vertical malocclusions.
Keywords: Biomechanics; Force control; Leveling archwire; Orthodontic simulation; Orthodontic tooth movement; Robotics.