Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Randomized Controlled Trial
. 2023 Sep;42(4):342-346.
doi: 10.1016/j.hansur.2023.05.009. Epub 2023 Jun 21.

Validation of a simplified evaluation grid applied to the microsurgery on inert material via smartphone

Affiliations
Randomized Controlled Trial

Validation of a simplified evaluation grid applied to the microsurgery on inert material via smartphone

Alexandre Thibault et al. Hand Surg Rehabil. 2023 Sep.

Abstract

Objectives: Microsurgery is usually performed in experimental research models and clinical surgery. It requires meticulous technical skills and continuous training. Inert materials such as Shirataki noodles are readily available and low-cost consumables regularly used for practice. The objective of this study was to evaluate the repeatability and reproducibility of a simplified evaluation grid of suture on inert material (Shirataki Konnyaku noodle) under smartphone magnification.

Material and methods: Ten students performed end-to-end suture on inert material with magnification via their smartphone. Each suture was filmed, and the videos were randomized. Each student was evaluated on each video three times over three consecutive days, using a simplified evaluation grid. Intra- and inter-observer agreement was evaluated on Concordance Correlation Coefficients. Values were assessed on Pearson's correlation coefficient.

Results: Intra-observer correlation was weak for 2 items (0.288 and 0.246) and moderate for the other 2 (0.419 and 0.529). Inter-observer correlation was weak for 3 items (0.344, 0.358, and 0.276) and close to zero for the other (0.034).

Conclusion: This simplified evaluation grid for microsurgery training on inert material via smartphone was poorly repeatable and reproducible. Loss of certain items in the grid due to the use of inert material probably impaired relevance.

Keywords: Anastomose microvasculaire; Inter- and intra-observer; Inter- intra-observateur; Konnyaku Shirataki; Microchirurgie; Microsurgery; Microvascular anastomosis; Smartphone.