. 1986;9(1-2):113-20.
doi: 10.1007/BF01743061.

The role of the pulsatile pressure variations in intracranial pressure monitoring

The role of the pulsatile pressure variations in intracranial pressure monitoring

C J Avezaat et al. Neurosurg Rev. 1986.

Abstract

The magnitude of the pulsatile intracranial pressure variations (CSF pulse pressure) is determined by the elastance of the craniospinal system and by the magnitude of the pulsatile variations in cerebral blood volume (CBV). The pulsatile change in CBV is, among other factors, determined by the compliance of the cerebral vascular bed which, in its turn, is dependent on the cerebral vasomotor tone. This concept has led the authors to devise a method for the assessment of both the elastance and the state of the cerebral vasomotor tone based on the relationship between CSF pulse pressure and intracranial pressure. This relationship was found to be of a linear nature both in clinical patients and in experimental animals. A significant, positive correlation was found between the slope of this relationship and the value of the craniospinal volume-pressure relationship: the elastance coefficient. During elevation of the intracranial pressure a breakpoint was observed in the relationship between CSF pulse pressure and the intracranial pressure above which the pulse pressure increased more rapidly. The elastance remained constant above this breakpoint. The same phenomenon was observed during plateau waves in clinical patients. Induced changes in systemic arterial pressure produced opposite effects on CSF pulse pressure and elastance coefficient. In these cases the discrepancy between pulse pressure and elastance was attributed to the pulsatile changes in CBV and this could be verified by means of electromagnetic flowmetry. The advantage of this method is that all the information is contained within the intracranial pressure signal itself, from which it can be extracted by simple means without the use of invasive tests.

PubMed Disclaimer

References

    1. J Neurosurg. 1975 Nov;43(5):523-34 - PubMed
    1. Acta Neurol Scand. 1973;49(5):557-74 - PubMed
    1. J Neurol Neurosurg Psychiatry. 1980 Mar;43(3):222-34 - PubMed
    1. Neurology. 1965 Jul;15:622-41 - PubMed
    1. Arch Neurol. 1973 Apr;28(4):265-9 - PubMed