In vitro comparative study of red blood cell and VWF damage on 3D printing biomaterials under different blood-contacting conditions
- PMID: 37417741
- DOI: 10.1177/09544119231186474
In vitro comparative study of red blood cell and VWF damage on 3D printing biomaterials under different blood-contacting conditions
Abstract
Mechanical circulatory support devices (MCSDs) are often associated with hemocompatible complications such as hemolysis and gastrointestinal bleeding when treating patients with end-stage heart failure. Shear stress and exposure time have been identified as the two most important mechanical factors causing blood damage. However, the materials of MCSDs may also induce blood damage when contacting with blood. In this study, the red blood cell and von Willebrand Factor (VWF) damage caused by four 3D printing biomaterials were investigated, including acrylic, PCISO, Somos EvoLVe 128, and stainless steel. A roller pump circulation experimental platform and a rotor blood-shearing experimental platform were constructed to mimic static and dynamic blood-contacting conditions of materials in MCSDs, respectively. Free hemoglobin assay and VWF molecular weight analysis were performed on the experimental blood samples. It indicated that different 3D printing materials and technology could induce different levels of damage to red blood cells and VWF, with acrylic causing the least damage under both static and dynamic conditions. In addition, it was found that blood damage measured for the same material differed on the two platforms. Therefore, a combination of static and dynamic experiments should be used to comprehensively investigate the effects of blood damage caused by the material. It can provide a reference for the design and evaluation of materials in different components of MCSDs.
Keywords: 3D printing materials; Hemolysis; blood damage; mechanical circulatory support devices; von Willebrand Factor.
Similar articles
-
In vitro study of red blood cell and VWF damage in mechanical circulatory support devices based on blood-shearing platform.Proc Inst Mech Eng H. 2022 Jun;236(6):860-866. doi: 10.1177/09544119221088420. Epub 2022 Apr 3. Proc Inst Mech Eng H. 2022. PMID: 35369808
-
The influence of surface roughness on the damage of von Willebrand Factor under shear flow condition.Int J Artif Organs. 2022 Apr;45(4):412-420. doi: 10.1177/03913988211056961. Epub 2021 Nov 4. Int J Artif Organs. 2022. PMID: 34736346
-
Mathematical models for shear-induced blood damage based on vortex platform.Int J Artif Organs. 2022 Apr;45(4):397-403. doi: 10.1177/03913988211003587. Epub 2021 Mar 19. Int J Artif Organs. 2022. PMID: 33740880
-
Acquired von Willebrand syndrome associated with left ventricular assist device.Blood. 2016 Jun 23;127(25):3133-41. doi: 10.1182/blood-2015-10-636480. Epub 2016 May 3. Blood. 2016. PMID: 27143258 Free PMC article. Review.
-
von Willebrand factor disruption and continuous-flow circulatory devices.J Heart Lung Transplant. 2017 Nov;36(11):1155-1163. doi: 10.1016/j.healun.2017.06.004. Epub 2017 Jun 12. J Heart Lung Transplant. 2017. PMID: 28756118 Review.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous