. 2023 Aug 9;67(7):886-894.
doi: 10.1093/annweh/wxad039.

An approach to quantify ortho-phthalaldehyde contamination on work surfaces

Affiliations

An approach to quantify ortho-phthalaldehyde contamination on work surfaces

Caitlyn A Rogers et al. Ann Work Expo Health. .

Abstract

Ortho-phthalaldehyde (OPA) is used as a high-level disinfectant for reusable medical devices in healthcare settings. The ACGIH recently adopted a Threshold Limit Value-Surface Limit (TLV-SL; 25 µg/100 cm2) for OPA surface contamination to prevent induction of dermal and respiratory sensitization following dermal exposure. However, there is no current validated method to measure OPA surface contamination. This study aimed to develop a standardized approach for sample collection and quantitative determination of OPA from work surfaces for use in risk assessment practices. The reported method utilises readily available commercial wipes to collect surface samples coupled with direct detection of OPA via liquid chromatography time of flight mass spectrometry (LC-ToF-MS). This approach avoided complex derivatization steps commonly required for the analysis of aldehydes. Method evaluation was conducted in accordance with the Occupational Safety and Health Administration (OSHA) surface sampling guidelines. Overall recoveries of 25 µg/100 cm2 of OPA from stainless steel and glass surfaces were 70% and 72%, respectively. The reported LOD for this method was 1.1 µg/sample and the LOQ was 3.7 µg/sample. OPA remained stable on the sampling medium for up to 10 days, when stored at 4 °C. The method was demonstrated in a workplace surface assessment at a local hospital sterilising unit, successfully detecting OPA on work surfaces. This method is intended to supplement airborne exposure assessment and provide a quantitative assessment tool for potential dermal exposure. When used in conjunction with a thorough occupational hygiene program that includes hazard communication, engineering controls, and personal protective equipment, skin exposure and consequent sensitization risks in the workplace can be minimized.

Keywords: o-phthalaldehyde; Threshold Limit Value– Surface Limit (TLV– SL); surface sampling.

PubMed Disclaimer

Conflict of interest statement

None declared.

References

    1. Anderson SE, Meada BJ.. Potential health effects associated with dermal exposure to occupational chemicals. Environ Health Insights 2014:8(S1):51–62. doi: 10.4137/EHI.S15258. - DOI - PMC - PubMed
    1. Atiyeah K, Chitkara A, Achlatis S, Branski RC, Amin M.. Allergic reaction to ortho-phthalaldehyde following flexible laryngoscopy. Laryngoscope. 2015:125(10):2349–2352. doi: 10.1002/lary.25421. - DOI - PubMed
    1. Chen L, Eisenberg J, Mueller C, Burton N.. 2015. Evaluation of ortho-phthalaldehyde in eight healthcare facilities. Health Hazard Evaluation Report No. 2006-0238-3239. Cincinnati, USA: US Department of Health and Human Services, Center for Disease Control and Prevention, National Institute for Occupational Safety and Health.
    1. Cooper DE, White AA, Werkema AN, Auge BK.. Anaphylaxis following cystoscopy with equipment sterilized with cidex® opa (ortho-phthalaldehyde): a review of two cases. J Endourol. 2008:22(9):2181–2184. doi: 10.1089/end.2007.0358. - DOI - PubMed
    1. Fujita H, Ogawa M, Endo Y.. A case of occupational bronchial asthma and contact dermatitis caused by ortho-pthalaldehyde exposure in a medical worker. J Occup Health. 2006:48(6):413–416. doi: 10.1539/joh.48.413. - DOI - PubMed

Publication types