The mechanical failure of locally manufactured prosthetic feet from the Jaffna Jaipur Centre for Disability Rehabilitation (JJCDR), Sri Lanka

Affiliations

The mechanical failure of locally manufactured prosthetic feet from the Jaffna Jaipur Centre for Disability Rehabilitation (JJCDR), Sri Lanka

Gemma Ranson et al. Prosthet Orthot Int. .

Abstract

Background: The International Organization for Standardization (ISO) 10328 standards are used to verify the safety of lower limb prosthetic devices. However, the ISO 10328 tests are performed in sterile laboratory settings and do not account for environmental or sociocultural factors associated with prosthetic use. Most locally manufactured prosthetic feet from low-income and middle-income countries, which are safely used for years, do not meet these standards. In this study, we investigate the modes of wear patterns of naturally worn prosthetic feet from Sri Lanka.

Objective: To characterize wear patterns of locally manufactured prosthetic feet from low-income and middle-income countries.

Methods: Sixty-six replaced prosthetic feet from the Jaffna Jaipur Center of Disability and Rehabilitation were analyzed. Delamination between the keel and rest of the foot could not be detected with ultrasound. Sole wear pattern was quantified by photographing soles and dividing them into 200 rectangles and scoring the rectangle's wear from 1 to 9 (no to extreme wear). Homologous scores were averaged to create a contour map of prosthetic foot wear.

Results: The highest levels of wear occurred at the heel, end of the keel, and the perimeter of the prosthetic foot. All regions of the prosthetic feet had significantly different wear scores ( p < 0.005).

Conclusions: Locally manufactured solid ankle cushion heel feet display high levels of wear in localized areas of the sole of prosthetic feet, which can limit their life span. High levels of wear occur at the end of the keel, which would not be detectable in the ISO 10328 tests.

PubMed Disclaimer

References

    1. Amara J, Hendricks A. Healthcare issues of the Iraq and Afghan wars: Short- and long-term impacts on US veterans’ healthcare 1. Defense Secur Anal 2009; 25: 285–298.
    1. Zhang Y, Lazzarini PA, McPhail SM, et al. Global disability burdens of diabetes-related lower-extremity complications in 1990 and 2016. Diabetes Care 2020; 43: 964–974.
    1. Cho NH, Shaw JE, Karuranga S, et al. IDF Diabetes Atlas: Global estimates of diabetes prevalence for 2017 and projections for 2045. Diabetes Res Clin Pract 2018; 138: 271–281.
    1. GHDx. GBD. Results Tool; 2019. Available at: http://ghdx.healthdata.org/gbd-results-tool?params=gbd-api-2019-permalin... .
    1. ISO. BSI Standards Publication Prosthetics—Structural Testing of Lower-Limb Prostheses—Requirements and Test Methods (ISO 10328:2016). Geneva, Switzerland: International Organization for Standardization; 2016.

LinkOut - more resources