Comparison between wavefront-derived refraction and auto-refraction
- PMID: 37459941
- DOI: 10.1016/j.pdpdt.2023.103712
Comparison between wavefront-derived refraction and auto-refraction
Abstract
Background: This study aimed to compare objective refractive errors and keratometry measurements obtained using the Nidek OPD-Scan II aberrometer/topographer and Topcon KR 8900 autorefractokeratometer.
Methods: The right eye medical records of 176 patients aged 18-35 years who were admitted to our clinic as refractive surgery candidates were tested for refractive status and keratometry measurements with a Nidek OPD-Scan II aberrometer/topographer and a standard table-top autorefractokeratometer (Topcon KR 8900) before and after the induction of cycloplegia. Patients who had undergone any eye surgery and had hereditary, ectatic, or acquired corneal pathology were excluded. Refractive data were compared as spheres, cylinders, spherical equivalents, and power vectors before and after the induction of cycloplegia. Flat and steep keratometry (K1-K2) readings were recorded in diopters (D) and axis degrees, respectively, for each eye.
Results: The spherical, cylindrical, spherical equivalence, J0-J45 vector values and K1-K2 readings (D, axis) between the two devices were statistically significant before and after the induction of cycloplegia (p<0.05). Bland-Altman analysis identified mean differences (95%CI of limits of agreement) of 0.77 (-0,57 to 2,11) in sphere, 0.74 (-0,54 to 2,01) in spherical equivalent, -0,07 (-0,41 to 0,26) in J0 vector, 0,06 (-0,31 to 0,43) in J45 vector, -0,16 (-0,66 to 0,33) in K1, -0,23 (-0,79 to 0,33) in K2 values before induction of cycloplegia.
Conclusion: The refractive and keratometry results of the Nidek OPD Scan II system and Topcon KR 8900 standard table-top autorefractokeratometer are not interchangeable in healthy adult population before and after induction of cycloplegia.
Keywords: Autorefraction; Cycloplegia; Ocular refraction; Topography; Wavefront aberration.