Accurate and efficient rapid acquisition early post-injection stress-first CZT SPECT myocardial perfusion imaging with tetrofosmin and attenuation correction
- PMID: 37464251
- DOI: 10.1007/s12350-023-03336-x
Accurate and efficient rapid acquisition early post-injection stress-first CZT SPECT myocardial perfusion imaging with tetrofosmin and attenuation correction
Abstract
Introduction: Myocardial perfusion imaging (MPI) protocols have not changed significantly despite advances in instrumentation and software. We compared an early post-injection, stress-first SPECT protocol to standard delayed imaging.
Methods: 95 patients referred for SPECT MPI were imaged upright and supine on a Spectrum Dynamics D-SPECT CZT system with CT attenuation correction. Patients received injection of 99mTc tetrofosmin at peak of regadenoson stress and were imaged. Early post-stress (mean 17 ± 2 minutes) and Standard 1-h delay (mean 61 ± 13 min). Three blinded readers evaluated images for overall interpretation, perceived need for rest imaging, image quality, and reader confidence. Laboratory efficiency was also evaluated.
Results: Blinded readers had the same response for the need for rest in 77.9% of studies. Studies also had the same interpretation in 89.5% of studies. Reader confidence was high (86.0% (Early) and 90.3% (Standard p = 0.52. Image quality was good or excellent in 87.4% Early vs 96.8% Standard (p = 0.09). Time between patient check-in and end of stress imaging was 104 ± (Standard) to 60 ± 18 minutes (Early) (p < 0.001).
Conclusion: Early post-injection stress-only imaging using CZT SPECT/CT appears promising with Tc-99m tetrofosmin with similar image quality, reader confidence, diagnosis, and need for a rest scan.
Keywords: CZT instrumentation; Early SPECT imaging; Tetrofosmin.
© 2023. The Author(s) under exclusive licence to American Society of Nuclear Cardiology.
Comment in
-
Time-efficient cardiac imaging: early post-injection stress-first protocol myocardial perfusion SPECT with a CZT camera.J Nucl Cardiol. 2023 Dec;30(6):2655-2657. doi: 10.1007/s12350-023-03369-2. Epub 2023 Aug 31. J Nucl Cardiol. 2023. PMID: 37653261 No abstract available.
Similar articles
-
Real-time respiratory triggered SPECT myocardial perfusion imaging using CZT technology: impact of respiratory phase matching between SPECT and low-dose CT for attenuation correction.Eur Heart J Cardiovasc Imaging. 2017 Jan;18(1):31-38. doi: 10.1093/ehjci/jew031. Epub 2016 Mar 16. Eur Heart J Cardiovasc Imaging. 2017. PMID: 26985074
-
Diagnostic accuracy of cadmium-zinc-telluride-based myocardial perfusion SPECT: impact of attenuation correction using a co-registered external computed tomography.Eur Heart J Cardiovasc Imaging. 2016 Sep;17(9):1036-43. doi: 10.1093/ehjci/jev312. Epub 2015 Nov 30. Eur Heart J Cardiovasc Imaging. 2016. PMID: 26628617
-
Comparison between N13NH3-PET and 99mTc-Tetrofosmin-CZT SPECT in the evaluation of absolute myocardial blood flow and flow reserve.J Nucl Cardiol. 2021 Oct;28(5):1906-1918. doi: 10.1007/s12350-019-01939-x. Epub 2019 Nov 14. J Nucl Cardiol. 2021. PMID: 31728817
-
A performance comparison of novel cadmium-zinc-telluride camera and conventional SPECT/CT using anthropomorphic torso phantom and water bags to simulate soft tissue and breast attenuation.Ann Nucl Med. 2015 May;29(4):342-50. doi: 10.1007/s12149-015-0952-z. Epub 2015 Jan 28. Ann Nucl Med. 2015. PMID: 25628019
-
Myocardial blood flow evaluation with dynamic cadmium-zinc-telluride single-photon emission computed tomography: Bright and dark sides.Diagn Interv Imaging. 2023 Jul-Aug;104(7-8):323-329. doi: 10.1016/j.diii.2023.02.001. Epub 2023 Feb 14. Diagn Interv Imaging. 2023. PMID: 36797156 Review.
Cited by
-
Time-efficient cardiac imaging: early post-injection stress-first protocol myocardial perfusion SPECT with a CZT camera.J Nucl Cardiol. 2023 Dec;30(6):2655-2657. doi: 10.1007/s12350-023-03369-2. Epub 2023 Aug 31. J Nucl Cardiol. 2023. PMID: 37653261 No abstract available.
References
-
- Thompson RC, Phillips LM, Dilsizian V, et al. Update on guidance and best practices for nuclear cardiology laboratories during the coronavirus disease 2019 (COVID-19) pandemic: emphasis on transition to chronic endemic state. An information statement from ASNC, IAEA, and SNMMI. J Nucl Cardiol 2022;29:1‐6. - DOI
-
- Skali H, Murthy VL, Paez D, et al. Guidance and best practices for reestablishment of non-emergent care in nuclear cardiology laboratories during the coronavirus disease 2019 (COVID-19) pandemic: an information statement from ASNC, IAEA, and SNMMI: endorsed by the Infectious Diseases Society of America. J Nucl Cardiol 2020;27:1855‐62. - DOI - PubMed - PMC
-
- Skulstad H, Cosyns B, Popescu BA, et al. COVID-19 pandemic and cardiac imaging: EACVI recommendations on precautions, indications, prioritization, and protection for patients and healthcare personnel. Eur Heart J 2020;21:592‐8.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources