Preclinical evaluation of a novel single-shot pulsed field ablation system for pulmonary vein and atrial ablation
- PMID: 37464948
- DOI: 10.1111/jce.16010
Preclinical evaluation of a novel single-shot pulsed field ablation system for pulmonary vein and atrial ablation
Abstract
Introduction: Pulsed field ablation (PFA) is a nonthermal ablative strategy that achieves cell death via electroporation. Herein, we investigated the preclinical safety and efficacy of PFA using two novel 8-French, 16-electrode spiral PFA/mapping catheters (ElePulse; CRC EP, Inc.).
Methods: Bipolar PFA (>1.8 kV) was performed using 30 s, single-shot, QRS-gated applications. Altogether, 94 atrial structures were ablated in 23 swine, one canine, and one ovine, including right and left atria and atrial appendages, pulmonary veins, and superior and inferior (IVC) vena cavae. We also examined the impact of PFA on the phrenic nerve (14 swine) and on a deviated esophagus after delivery of PFA from inside the IVC (five swine).
Results: All applications were single-shot without catheter repositioning. Minimal microbubbling was observed without significant skeletal muscle twitching/activation (mean acceleration: 0.05 m/s2 ). There was a marked reduction in post-PFA versus pre-PFA atrial electrogram amplitude (0.17 ± 0.21 vs. 1.18 ± 1.08 mV; p < .0001). Lesion durability was demonstrated up to 3 months in all targeted tissues. Histologically, lesions were contiguous and transmural, except in the atrial appendage, and without any thermal effects. Magnetic resonance, gross, and histologic examinations of the brain, rete mirabile, and kidneys revealed no thromboembolism. No acute/long-term phrenic nerve dysfunction was encountered. Although within 2 h of ablation, histologic examinations of the esophagus revealed acute PFA-related changes in the muscular layer, these completely resolved by 21 ± 5 days.
Conclusion: A novel, single-shot, spiral PFA system is capable of safely creating large, durable atrial lesions without significant adverse effects on the phrenic nerve or the esophagus.
Keywords: catheter ablation; electroporation; esophagus; phrenic nerve; pulmonary vein isolation; pulsed field ablation.
© 2023 Wiley Periodicals LLC.
Comment in
-
A novel single shot pulsed field ablation catheter: Where does it stand and where do we go from here?J Cardiovasc Electrophysiol. 2023 Nov;34(11):2213-2215. doi: 10.1111/jce.16033. Epub 2023 Aug 9. J Cardiovasc Electrophysiol. 2023. PMID: 37554100 No abstract available.
References
REFERENCES
-
- Maor E, Sugrue A, Witt C, et al. Pulsed electric fields for cardiac ablation and beyond: a state-of-the-art review. Heart Rhythm. 2019;16:1112-1120.
-
- Mali B, Jarm T, Snoj M, Sersa G, Miklavcic D. Antitumor effectiveness of electrochemotherapy: a systematic review and meta-analysis. Eur J Surg Oncol. 2013;39:4-16.
-
- Heller R, Heller LC. Gene electrotransfer clinical trials. Adv Genet. 2015;89:235-262.
-
- Reddy VY, Koruth J, Jais P, et al. Ablation of atrial fibrillation with pulsed electric fields. JACC Clin Electrophysiol. 2018;4:987-995.
-
- DeSimone CV, Kapa S, Asirvatham SJ. Electroporation: past and future of catheter ablation. Circ Arrhythm Electrophysiol. 2014;7:573-575.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical