Case Reports
. 2023 Sep;102(3):495-498.
doi: 10.1002/ccd.30768. Epub 2023 Jul 18.

Management of a broken guiding catheter tip: Cut and fix technique

Affiliations
Case Reports

Management of a broken guiding catheter tip: Cut and fix technique

Navreet Singh et al. Catheter Cardiovasc Interv. 2023 Sep.

Abstract

A 49-year-old male presented with class III exertional angina, 1 year after angioplasty of the left anterior descending artery (LAD) and right coronary artery. Coronary angiogram revealed 90% in-stent restenosis (ISR) in mid-LAD with angiographic impression of stent fracture. Optical coherence tomographic evaluation of mid-LAD ISR showed a distinct 3 mm long "eclipse sign" indicating embolized, broken guiding catheter tip as a cause of ISR, which was confirmed on reviewing 1-year-old angiographic images. This was managed with "cut and fix technique" using cutting balloon and another drug-eluting stent. Optical coherence tomographic at 9 months showed well endothelialized stent with a thin layer of neo-intimal hyperplasia over the sandwiched broken guiding tip.

Keywords: broken catheter tip; cut-and-fix; eclipse sign; optical coherence tomography.

PubMed Disclaimer

References

REFERENCES

    1. Trehan V, Mukhopadhyay S, Yusuf J, C. Ramgasetty U, Mukherjee S, Arora R. Intracoronary fracture and embolization of a coronary angioplasty balloon catheter: retrieval by a simple technique. Catheter Cardiovasc Interv. 2003;58:473-477.
    1. Manjunath CN, Bhat P, Panneerselvam A. Revascularization after removal of broken catheter from left circumflex coronary artery. Cardiovasc Med. 2011;14(7-8):222-224.
    1. Lalani K, Shroff J, Parikh P. If a broken catheter tip cannot be retrieved from the coronary artery, ‘fix’ it! Coron Artery Dis. 2022;33:692-694.
    1. Gupta A, Shrivastava A, Vijayvergiya R, et al. Optical coherence tomography: an eye into the coronary artery. Front Cardiovasc Med. 2022;9:854554. doi:10.3389/fcvm.2022.854554
    1. Gupta A, Chhikara S, Datta R, Vijayvergiya R. Everolimus-eluting stent causing coronary artery aneurysm in 7 days: 3D-OCT findings and management. J Invasive Cardiol. 2020;32(11):301.