The intrinsic signal-to-noise ratio in NMR imaging
- PMID: 3747821
- DOI: 10.1002/mrm.1910030413
The intrinsic signal-to-noise ratio in NMR imaging
Abstract
The fundamental limit for NMR imaging is set by an intrinsic signal-to-noise ratio (SNR) for a particular combination of rf antenna and imaging subjects. The intrinsic SNR is the signal from a small volume of material in the sample competing with electrical noise from thermally generated, random noise currents in the sample. The intrinsic SNR has been measured for a number of antenna-body section combinations at several different values of the static magnetic field and is proportional to B0. We have applied the intrinsic and system SNR to predict image SNR and have found satisfactory agreement with measurements on images. The relationship between SNR and pixel size is quite different in NMR than it is with imaging modalities using ionizing radiation, and indicates that the initial choice of pixel size is crucial in NMR. The analog of "contrast-detail-dose" plots for ionizing radiation imaging modalities is the "contrast-detail-time" plot in NMR, which should prove useful in choosing a suitable pixel array to visualize a particular anatomical detail for a given NMR receiving antenna.
Similar articles
-
An unbiased signal-to-noise ratio measure for magnetic resonance images.Med Phys. 1993 Jul-Aug;20(4):1077-8. doi: 10.1118/1.597004. Med Phys. 1993. PMID: 8413015
-
On the Contribution of Curl-Free Current Patterns to the Ultimate Intrinsic Signal-to-Noise Ratio at Ultra-High Field Strength.NMR Biomed. 2017 May;30(5). doi: 10.1002/nbm.3691. Epub 2017 Feb 10. NMR Biomed. 2017. PMID: 28186652
-
Computation of the signal-to-noise ratio of high-frequency magnetic resonance imagers.IEEE Trans Biomed Eng. 2000 Nov;47(11):1525-33. doi: 10.1109/10.880105. IEEE Trans Biomed Eng. 2000. PMID: 11077747
-
Signal-to-noise measurements in magnitude images from NMR phased arrays.Magn Reson Med. 1997 Nov;38(5):852-7. doi: 10.1002/mrm.1910380524. Magn Reson Med. 1997. PMID: 9358462 Free PMC article.
-
Magnetic resonance imaging: present and future applications.Can Med Assoc J. 1985 Apr 1;132(7):765-77. Can Med Assoc J. 1985. PMID: 3884120 Free PMC article. Review.
Cited by
-
Toward imaging the body at 10.5 tesla.Magn Reson Med. 2017 Jan;77(1):434-443. doi: 10.1002/mrm.26487. Epub 2016 Oct 21. Magn Reson Med. 2017. PMID: 27770469 Free PMC article.
-
Sodium MRI in human heart: a review.NMR Biomed. 2016 Feb;29(2):187-96. doi: 10.1002/nbm.3265. Epub 2015 Feb 12. NMR Biomed. 2016. PMID: 25683054 Free PMC article. Review.
-
Sub-millimeter T2 weighted fMRI at 7 T: comparison of 3D-GRASE and 2D SE-EPI.Front Neurosci. 2015 May 5;9:163. doi: 10.3389/fnins.2015.00163. eCollection 2015. Front Neurosci. 2015. PMID: 25999810 Free PMC article.
-
Fluid attenuated inversion recovery (FLAIR) imaging of the normal brain: comparisons between under the conditions of 3.0 Tesla and 1.5 Tesla.Korean J Radiol. 2010 Jan-Feb;11(1):19-24. doi: 10.3348/kjr.2010.11.1.19. Epub 2009 Dec 28. Korean J Radiol. 2010. PMID: 20046491 Free PMC article.
-
In vivo human head MRI at 10.5T: A radiofrequency safety study and preliminary imaging results.Magn Reson Med. 2020 Jul;84(1):484-496. doi: 10.1002/mrm.28093. Epub 2019 Nov 21. Magn Reson Med. 2020. PMID: 31751499 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources