Review
. 2023 Jul 25;18(1):212.
doi: 10.1186/s13023-023-02801-7.

Gaps in the evidence underpinning high-risk medical devices in Europe at market entry, and potential solutions

Affiliations
Review

Gaps in the evidence underpinning high-risk medical devices in Europe at market entry, and potential solutions

Frank Hulstaert et al. Orphanet J Rare Dis. .

Abstract

Aim: To determine the level of evidence for innovative high-risk medical devices at market entry.

Methods: We reviewed all Belgian healthcare payer (RIZIV-INAMI) assessor reports on novel implants or invasive medical devices (n = 18, Class IIb-III) available between 2018 to mid-2019 on applications submitted for inclusion on their reimbursement list. We also conducted a review of the literature on evidence gaps and an analysis of relevant legal and ethical frameworks within the European context.

Findings: Conformity assessment of medical devices is based on performance, safety, and an acceptable risk-benefit balance. Information submitted for obtaining CE marking is confidential and legally protected, limiting access to clinical evidence. Seven out of the 18 RIZIV-INAMI assessor reports (39%) included a randomized controlled trial (RCT) using the novel device, whilst 2 applications (11%) referred to an RCT that used a different device. The population included was inappropriate or unclear for 3 devices (17%). Only half of the applications presented evidence on quality of life or functioning and 2 (11%) presented overall survival data. Four applications (22%) included no data beyond twelve months. The findings from the literature demonstrated similar problems with the study design and the clinical evidence.

Discussion and conclusions: CE marking does not indicate that a device is effective, only that it complies with the law. The lack of transparency hampers evidence-based decision making. Despite greater emphasis on clinical benefit for the patient, the provisions of the European Medical Device Regulation (MDR) are not yet fully aligned with international ethical standards for clinical research. The MDR fails to address key issues, such as the lack of access to data submitted for CE marking and a failure to require evidence of clinical effectiveness. Indeed, a first report shows no improvement in the clinical evidence for implantable devices generated under the MDR. Thus, patients may continue to be exposed to ineffective or unsafe novel devices. The Health Technology Assessment Regulation plans for Joint Scientific Consultations for specific high-risk devices before companies begin their pivotal clinical investigations. The demanded comparative evidence should facilitate payer decisions. Nevertheless, there is also a need for legislation requiring comparative RCTs assessing patient-relevant outcomes for high-risk devices to ensure implementation, including development and implementation of common specifications for study designs.

Keywords: CE marking; Comparative evidence; Comparator; Declaration of Helsinki; Evidence gaps; Medical devices; Reimbursement; Study design.

PubMed Disclaimer

Conflict of interest statement

All authors have no competing interests.

Figures

Fig. 1
Fig. 1
Evidence requirements necessary for patient access to novel health technologies

References

    1. European Economic and Social Committee. Opinion - Amended proposal for a Regulation of the European Parliament and of the Council on health technology assessment and amending Directive 2011/24/EU 2021 [cited 14/9/2021]. Available from: https://eur-lex.europa.eu/legal-content/EN/TXT/PDF/?uri=CONSIL:ST_8330_2....
    1. Hulstaert F, Pouppez C, Primus-de Jong C, Harkin K, Neyt M. Evidence gaps for drugs and medical devices at market entry in Europe and potential solutions. Health Services Research (HSR. Brussel: Belgian Health Care Knowledge Centre (KCE); 2021.
    1. Ergina PL, Cook JA, Blazeby JM, Boutron I, Clavien PA, Reeves BC, et al. Challenges in evaluating surgical innovation. Lancet. 2009;374(9695):1097–104. doi: 10.1016/S0140-6736(09)61086-2. - DOI - PMC - PubMed
    1. Barkun JS, Aronson JK, Feldman LS, Maddern GJ, Strasberg SM, Balliol C, et al. Evaluation and stages of surgical innovations. Lancet. 2009;374(9695):1089–96. doi: 10.1016/S0140-6736(09)61083-7. - DOI - PubMed
    1. Sedrakyan A, Marinac-Dabic D, Normand SL, Mushlin A, Gross T. A framework for evidence evaluation and methodological issues in implantable device studies. Med Care. 2010;48(6 Suppl):121–8. doi: 10.1097/MLR.0b013e3181d991c4. - DOI - PubMed