Survival From Revision Surgery for New Rigicon Infla10 Three-piece Inflatable Penile Prosthesis Is Comparable to Preceding Devices

Affiliations
Free article

Survival From Revision Surgery for New Rigicon Infla10 Three-piece Inflatable Penile Prosthesis Is Comparable to Preceding Devices

Steven K Wilson et al. Urology. 2023 Oct.
Free article

Abstract

Objective: To report the incidence of the reoperation surgeries of nearly all the Rigicon Infla10 implants performed since device introduction in 1/2019. Inflatable penile prosthesis has some of the highest survival from revision surgery of any medical device implanted in humans [1]. We expand on previous Rigicon Infla10 research, adding more patients and increasing follow-up duration [2].

Materials and methods: 535 patients had Rigicon Infla10 devices implanted from 1/2019 to 8/2022. 103 surgeons from 26 centers in 15 countries participated in the study. Patient information forms were analyzed from virtually all implantations. Explantation or revision surgery for mechanical failure, infection, other medical reasons, and patient dissatisfaction were cataloged. SPSS 25.0 (IBM) was used for the statistical analysis of Kaplan Meier survival statistics.

Results: Mean follow-up was 24.2months (7-43months). Mean patient age was 56years. Reoperation was necessary for 3.5% of subjects. Revision for mechanical failure occurred in 2.24% (12/535). The rate of explant for patient dissatisfaction was 0.56% (3/535). Revision for component out of place was 0.37% (2/535) with an infection rate and unsuccessful Peyronie's correction being 0.19% (1/535). Survival from requiring another corrective surgery at 1, 2, and 3years was 96.4%, 95.0%, and 94.0%, respectively. These initial survival rates compare favorably to devices currently available, which have been repeatedly enhanced to improve reliability.

Conclusion: In its first 2-3years of availability, The Rigicon Infla10 inflatable penile prosthesis shows freedom from revision comparable to existing enhanced devices that have been on the market for decades.

PubMed Disclaimer

Conflict of interest statement

Declaration of Competing Interest The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: SW: International Medical Device, Rigicon, Uramix. HM: None. BK: None. GT: Rigicon. CL: Coloplast. MR: None. AC: None. MH: None. MG: Coloplast.