Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2023 Jul 23:22:101170.
doi: 10.1016/j.artd.2023.101170. eCollection 2023 Aug.

Magnetic Resonance Safety Evaluation of a Novel Alumina Matrix Composite Ceramic Knee and Image Artifact Comparison to a Metal Knee Implant of Analogous Design

Affiliations

Magnetic Resonance Safety Evaluation of a Novel Alumina Matrix Composite Ceramic Knee and Image Artifact Comparison to a Metal Knee Implant of Analogous Design

Yvonne Mödinger et al. Arthroplast Today. .

Abstract

Background: Image artifacts caused by metal knee implants in 1.5T and 3T magnetic resonance imaging (MRI) systems complicate imaging-based diagnosis of the peri-implant region after total knee arthroplasty. Alternatively, metal-free knee prostheses could effectively minimize MRI safety hazards and offer the potential for higher quality diagnostic images.

Methods: A novel knee arthroplasty device composed of BIOLOX delta, an alumina matrix composite (AMC) ceramic, was tested in a magnetic resonance (MR) environment. American Society for Testing and Materials test methods were used for evaluating magnetically induced displacement force, magnetically induced torque, radiofrequency-induced heating, and MR image artifact.

Results: Magnetically induced displacement force and magnetically induced torque results of the AMC ceramic knee indicated that these effects do not pose a known risk in a clinical MR environment, as assessed in a 3T magnetic field. Moreover, minimal radiofrequency-induced heating of the device was observed. In addition, the AMC ceramic knee demonstrated minimal MR image artifacts (7 mm) in comparison to a cobalt-chromium knee (88 mm). The extremely low magnetic susceptibility of AMC (2 ppm) underlines that it is a nonmetallic and nonmagnetic material well suited for the manufacturing of MR Safe orthopaedic implants.

Conclusions: The AMC ceramic knee is a novel metal-free total knee arthroplasty device that can be regarded as MR Safe, as suggested by the absence of hazards from the exposure of this implant to a MR environment. The AMC ceramic knee presents the advantage of being scanned with superior imaging results in 3T MRI systems compared to alternative metal implants on the market.

Keywords: Alumina matrix composite; Ceramic; Image artifact; Knee implant; Magnetic resonance imaging; Total knee arthroplasty.

PubMed Disclaimer

Figures

Figure 1
Figure 1
RF-induced temperature rises for the AMC ceramic knee after 15 minutes of scanning in a 3T MRI system, normalized to the initial temperature.
Figure 2
Figure 2
AMC ceramic knee with a spin echo pulse sequence in a 3T MRI system. Worst-case image (left), background image (middle), resulting image artifact (right). Measurements are in millimeters.
Figure 3
Figure 3
AMC ceramic knee with a gradient echo pulse sequence in a 3T MRI system. Worst-case image (left), background image (middle), resulting image artifact (right). Measurements are in millimeters.
Figure 4
Figure 4
CoCr knee with a spin echo pulse sequence in a 3T MRI system. Worst-case image (left), background image (middle), resulting image artifact (right). Measurements are in millimeters.
Figure 5
Figure 5
CoCr knee with a gradient echo pulse sequence in a 3T MRI system. Worst-case image (left), background image (middle), resulting image artifact (right). Measurements are in millimeters.
Fig. S1
Fig. S1
Slices through the empty gel phantom of electric field in V/m (top) and SAR in W/kg (bottom) for simulated RF application in a 3T/123 MHz MRI system. Computational analysis of the gelled saline phantom showed that the left side of the gelled saline phantom is associated with a greater electrical field. Thus, the test article was positioned on the far-left side of the gelled saline phantom (not shown).
Fig. S2
Fig. S2
Fiber Optic Thermometry Probe locations during RF-induced heating evaluation on a 3T MRI system (Discovery MR750 [GE Healthcare]). Three probes were placed on the implant. The fourth probe served as a reference and was located on the opposite side of the gel phantom.

References

    1. Asif M.I. Characterisation and biological impact of wear particles from composite ceramic hip replacements. School of Mechanical Engineering, The University of Leeds; 2018. PhD Thesis.
    1. Cunningham B.W., Hallab N.J., Hu N., McAfee P.C. Epidural application of spinal instrumentation particulate wear debris: a comprehensive evaluation of neurotoxicity using an in vivo animal model. J Neurosurg Spine. 2013;19:336–350. - PubMed
    1. Zietz C., Bergschmidt P., Lange R., Mittelmeier W., Bader R. Third-body abrasive wear of tibial polyethylene inserts combined with metallic and ceramic femoral components in a knee simulator study. Int J Artif Organs. 2013;36:47–55. - PubMed
    1. Maccauro G., Cittadini A., Magnani G., et al. In vivo characterization of zirconia toughened alumina material: a comparative animal study. Int J Immunopathol Pharmacol. 2010;23:841–846. - PubMed
    1. Maccauro G., Bianchino G., Sangiorgi S., Magnani G., Marotta D., Manicone P.F., et al. Development of a new zirconia-toughened alumina: promising mechanical roperties and absence of in vitro carcinogenicity. Int J Immunopathol Pharmacol. 2009;22:773–779. - PubMed

LinkOut - more resources