The measurement of Central Corneal Thickness

Affiliations

The measurement of Central Corneal Thickness

Georgiana Camburu et al. Rom J Ophthalmol. 2023 Apr-Jun.

Abstract

We aimed to compare five different devices that measure Central Corneal Thickness. The Central Corneal Thickness (CCT) is an important parameter in ophthalmology. It is involved in the management of various eye conditions such as: glaucoma, keratoconus, contact lens wearing, corneal dystrophies, refractive surgery and keratoplasty. We measured the CCT using OCT, Topographer (TOPO), Ultrasonography Pachymeter (US), Specular Microscope (MS), and Non-contact Tonometer (TONO). In the analysis of the data collected from 59 patients we found the following mean values: US - 554.51 ± 29.849 μm, OCT - 548.73 ± 31.080 μm, TOPO - 553.76 ± 29.845 μm, MS - 564 ± 32.637 μm, and TONO - 538.9 ± 35.657 μm. Our results confirmed the strong correlation between techniques. Abbreviations: OCT = Optical Coherence Tomography, CCT = Central Corneal Thickness, TOPO = Topographer, US = Ultrasonography Pachymeter, MS = Specular Microscope, TONO = Non-contact Tonometer.

Keywords: Central Corneal Thickness; OCT; Pachymetry; Specular Microscopy.

PubMed Disclaimer

Figures

Graph 1
Graph 1
US-OCT
Graph 2
Graph 2
US-TOPO
Graph 3
Graph 3
US-MS
Graph 4
Graph 4
US-TONO
Graph 5
Graph 5
OCT-TOPO
Graph 6
Graph 6
OCT-MS
Graph 7
Graph 7
OCT-TONO
Graph 8
Graph 8
TOPO-MS
Graph 9
Graph 9
TOPO-TONO
Graph 10
Graph 10
MS-TONO

References

    1. Binnawi KH, Elzubeir H, Osman E, Abdu M, Abdu M. Central corneal thickness measurement using ultrasonic pachymeter, optical coherence tomography, and TMS-5 topographer. Oman J Ophthalmol. 2019 Jan-Apr;12(1):15–19. doi: 10.4103/ojo.OJO_96_2018. - PMC - PubMed
    1. Krachmer JH, Mannis MJ, Holland EJ. Cornea. Third Edition. St. Louis: Mosby Elsevier; 2011. p. 617.719-dc22.
    1. Garg A. Anterior and Posterior Segment OCT: Current Technology and Future Applications. First Ed. Jaypee Brothers Medical Publishers LTD; 2014.
    1. Garg A. Instant Clinical diagnosis in Ophthalmology Anterior Segment Diseases. Jaypee Brothers Medical Publishers LTD; 2010.
    1. Wang M, Kugler LJ. Atlas and Clinical Reference Guide for Corneal Topography. SLACK Incorporated; 2014.

LinkOut - more resources