Design and validation of an automated hydrostatic weighing system
- PMID: 3755792
Design and validation of an automated hydrostatic weighing system
Abstract
The purpose of this study was to design and evaluate the validity of an automated technique to assess body density using a computerized hydrostatic weighing system. An existing hydrostatic tank was modified and interfaced with a microcomputer equipped with an analog-to-digital converter. Software was designed to input variables, control the collection of data, calculate selected measurements, and provide a summary of the results of each session. Validity of the data obtained utilizing the automated hydrostatic weighing system was estimated by: evaluating the reliability of the transducer/computer interface to measure objects of known underwater weight; comparing the data against a criterion measure; and determining inter-session subject reliability. Values obtained from the automated system were found to be highly correlated with known underwater weights (r = 0.99, SEE = 0.0060 kg). Data concurrently obtained utilizing the automated system and a manual chart recorder were also found to be highly correlated (r = 0.99, SEE = 0.0606 kg). Inter-session subject reliability was determined utilizing data collected on subjects (N = 16) tested on two occasions approximately 24 h apart. Correlations revealed high relationships between measures of underwater weight (r = 0.99, SEE = 0.1399 kg) and body density (r = 0.98, SEE = 0.00244 g X cm-1). Results indicate that a computerized hydrostatic weighing system is a valid and reliable method for determining underwater weight.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical