Switching the Left and the Right Hearts: A Novel Bi-ventricle Mechanical Support Strategy with Spared Native Single-Ventricle
- PMID: 37635154
- DOI: 10.1007/s10439-023-03348-1
Switching the Left and the Right Hearts: A Novel Bi-ventricle Mechanical Support Strategy with Spared Native Single-Ventricle
Abstract
End-stage Fontan patients with single-ventricle (SV) circulation are often bridged-to-heart transplantation via mechanical circulatory support (MCS). Donor shortage and complexity of the SV physiology demand innovative MCS. In this paper, an out-of-the-box circulation concept, in which the left and right ventricles are switched with each other is introduced as a novel bi-ventricle MCS configuration for the "failing" Fontan patients. In the proposed configuration, the systemic circulation is maintained through a conventional mechanical ventricle assist device (VAD) while the venous circulation is delegated to the native SV. This approach spares the SV and puts it to a new use at the right-side providing the most-needed venous flow pulsatility to the failed Fontan circulation. To analyze its feasibility and performance, eight SV failure modes have been studied via an established multi-compartmental lumped parameter cardiovascular model (LPM). Here the LPM model is experimentally validated against the corresponding pulsatile mock-up flow loop measurements of a representative 15-year-old Fontan patient employing a clinically-approved VAD (Medtronic-HeartWare). The proposed surgical configuration maintained the healthy cardiac index (3-3.5 l/min/m2) and the normal mean systemic arterial pressure levels. For a failed SV with low ejection fraction (EF = 26%), representing a typical systemic Fontan failure, the proposed configuration enabled a ~ 28 mmHg amplitude in the venous/pulmonary waveforms and a 2 mmHg decrease in the central venous pressure (CVP) together with acceptable mean pulmonary artery pressures (17.5 mmHg). The pulmonary vascular resistance (PVR)-SV failure case provided a ~ 5 mmHg drop in the CVP, with venous/pulmonary pulsatility reaching to ~ 22 mmHg. For the high PVR failure case with a healthy SV (EF = 44%) pulmonary hypertension is likely to occur as expected. While this condition is routinely encountered during the heart transplantation and managed through pulmonary vasodilators a need for precise functional assessment of the spared failed-ventricle is recommended if utilized in the PVR failure mode. Comprehensive in vitro and in silico results encourage this novel concept as a low-cost, more physiological alternative to the conventional bi-ventricle MCS pending animal experiments.
Keywords: Cardiovascular circulation theory; Congenital heart surgery; Fontan circulation; Hemodynamics; Lumped parameter modelling; Mechanical circulatory support; Mock-up flow loops; Single-ventricle physiology; Ventricle assist devices.
© 2023. The Author(s) under exclusive licence to Biomedical Engineering Society.
Similar articles
-
Hemodynamic Effects of Ventricular Assist Device Implantation on Norwood, Glenn, and Fontan Circulation: A Simulation Study.Artif Organs. 2016 Jan;40(1):34-42. doi: 10.1111/aor.12591. Epub 2015 Nov 2. Artif Organs. 2016. PMID: 26526959
-
Coupling pediatric ventricle assist devices to the Fontan circulation: simulations with a lumped-parameter model.ASAIO J. 2005 Sep-Oct;51(5):618-28. doi: 10.1097/01.mat.0000176169.73987.0d. ASAIO J. 2005. PMID: 16322728
-
Use of Ventricular Assist Device in Univentricular Physiology: The Role of Lumped Parameter Models.Artif Organs. 2016 May;40(5):444-53. doi: 10.1111/aor.12583. Epub 2015 Oct 23. Artif Organs. 2016. PMID: 26494529
-
Failing Fontan cardiovascular support: Review.J Card Surg. 2022 Dec;37(12):5257-5261. doi: 10.1111/jocs.17094. Epub 2022 Nov 2. J Card Surg. 2022. PMID: 36321714 Free PMC article. Review.
-
Support with Single Ventricle-Ventricular Assist Device (sVAD) in Patients with Functionally Univentricular Circulation Prior to Fontan Operation.Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu. 2023;26:26-39. doi: 10.1053/j.pcsu.2022.12.002. Epub 2022 Dec 24. Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu. 2023. PMID: 36842796 Review.
References
-
- Bernstein, D., D. Naftel, C. Chin, L. J. Addonizio, P. Gamberg, E. D. Blume, D. Hsu, C. E. Canter, J. K. Kirklin, W. R. Morrow, Society for Pediatric Heart Transplantation. Outcome of listing for cardiac transplantation for failed Fontan: a multi-institutional study. Circulation. 114:273–280, 2006. - PubMed - DOI
-
- Cilliers, A., and M. Gewillig. Fontan procedure for univentricular hearts: have changes in design improved outcome? Cardiovasc. J. S. Afr. 13:111–116, 2002. - PubMed
MeSH terms
Grants and funding