Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2023 Nov;37(6):686-691.
doi: 10.1177/19458924231193520. Epub 2023 Aug 27.

Eustachian Tube Balloon Dilation: A Comprehensive Analysis of Adverse Events

Affiliations

Eustachian Tube Balloon Dilation: A Comprehensive Analysis of Adverse Events

Tyler B Merrill et al. Am J Rhinol Allergy. 2023 Nov.

Abstract

Background: Eustachian tube balloon dilation (ETBD) has been Food and Drug Administration (FDA) approved for refractory Eustachian tube dysfunction since 2016. While ETBD is generally seen as safe, the complication profile has not been well defined.

Objective: The objective of this study was to utilize the FDA manufacturer and user facility device experience (MAUDE) database to better assess adverse events (AE) related to ETBD.

Methods: This is a study of a multiinstitutional database maintained by the U.S. FDA. A database analysis was performed via the collaboration of multiple clinicians at tertiary referral centers. The FDA MAUDE database was queried for all medical device reports (MDR) related to ETBD devices from January 2012 to November 2022. Eighty-eight unique MDR were identified, 16 of which met inclusion criteria.

Results: Three MDRs were classified as device-related (18.8%); none resulted in an AE. Thirteen MDRs (81.3%) were patient-related; all were classified as AEs. The most common AE was postoperative subcutaneous emphysema (n = 6, 46.2%). Of the patients with subcutaneous emphysema, there was a wide range of severity. The most severe AE (n = 1, 6.3%) was postoperative stroke secondary to carotid artery dissection.

Conclusion: Though ETBD is generally seen as a safe procedure, there have been several concerning AEs reported to date. Increased awareness of ETBD complications can serve as a primer for improved patient education and counseling during the informed consent process and aid surgeons in clinical decision-making. Future studies with standardized reporting protocols are warranted to create a central registry for ETBD.

Keywords: MAUDE database; adverse event; balloon dilation; eustachian tube dysfunction; risk profile.

PubMed Disclaimer

Conflict of interest statement

Declaration of Conflicting InterestsThe author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

LinkOut - more resources