. 1986 Oct;161(1):175-81.
doi: 10.1148/radiology.161.1.3763863.

Radiation dose reduction in the evaluation of scoliosis: an application of digital radiography

Radiation dose reduction in the evaluation of scoliosis: an application of digital radiography

D C Kushner et al. Radiology. 1986 Oct.

Abstract

This report documents the clinical testing of scanning beam digital radiography as an imaging method in patients with scoliosis. This type of digital imaging requires a skin exposure of only 2.4 mR (0.619 microC/kg) per image, compared with the lowest possible posteroanterior screen-film exposure of 10 mR (2.58 microC/kg) at the chest and 60 mR (15.48 microC/kg) at the lumbar spine. Digital radiographic and screen-film images were obtained on multiple test objects and 273 patients. Scoliosis measurements using screen-film radiographs and digital radiographs were comparable to within a mean difference of 1 degrees at many different degrees of severity. The low-dose digital images were found to be useful and accurate for the detection and measurement of scoliosis after the first screen-film radiographs have excluded tumors and structural abnormalities.

PubMed Disclaimer

LinkOut - more resources