Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2023 Dec;30(6):2736-2749.
doi: 10.1007/s12350-023-03358-5. Epub 2023 Aug 28.

Detection and correction of patient motion in dynamic 15O-water PET MPI

Affiliations

Detection and correction of patient motion in dynamic 15O-water PET MPI

Nana L Christensen et al. J Nucl Cardiol. 2023 Dec.

Abstract

Background: Patient motion constitutes a limitation to 15O-water cardiac PET imaging. We examined the ability of image readers to detect and correct patient motion using simulated motion data and clinical patient scans.

Methods: Simulated data consisting of 16 motions applied to 10 motion-free scans were motion corrected using two approaches, pre-analysis and post-analysis for motion identification. Both approaches employed a manual frame-by-frame correction method. In addition, a clinical cohort was analyzed for assessment of prevalence and effect of motion and motion correction.

Results: Motion correction was performed on 94% (pre-analysis) and 64% (post-analysis) of the scans. Large motion artifacts were corrected in 91% (pre-analysis) and 74% (post-analysis) of scans. Artifacts in MBF were reduced in 56% (pre-analysis) and 58% (post-analysis) of the scans. The prevalence of motion in the clinical patient cohort (n = 762) was 10%. Motion correction altered exam interpretation in only 10 (1.3%) clinical patient exams.

Conclusion: Frame-by-frame motion correction after visual inspection is useful in reducing motion artifacts in cardiac 15O-water PET. Reviewing the initial results (parametric images and polar maps) as part of the motion correction process, reduced erroneous corrections in motion-free scans. In a large clinical cohort, the impact of motion correction was limited to few patients.

Keywords: PET; image analysis; myocardial blood flow; perfusion agents.

PubMed Disclaimer

Conflict of interest statement

N.L.C., J.N., S.M., M.A.M., L.C.G, and T.K. have nothing to disclose. M.L. and L.P.T. hold shares in MedTrace Pharma AS, Hørsholm, Denmark.

Figures

Figure 1
Figure 1
Frame-by-frame motion correction software. All 21 frames are displayed, showing an overlay of the left ventricular segmentation from the aQuant pre-correction analysis. Each frame can be manually shifted in the x-, y-, or z-direction
Figure 2
Figure 2
Heat maps of uncorrected (A, D, G) and motion corrected images using the pre-analysis (B, E, H) and post-analysis (C, F, I) approach. The X-axis represents each type of motion. The Y-axis represents the coronary territories (LAD, RCA, LCx, and global). Median relative deviation (%) in MBF from the original motion-free images in uncorrected (A) and motion-corrected (B) images. Maximum relative deviation (%) in MBF from the original motion-free images in uncorrected (D) and in motion-corrected images using the preanalysis approach (E) and post-analysis approach (F). Results of the Wilcoxon signed-rank test, showing level of significance (P > .05, P < .05 or P < .005) in MBF deviation from the original motion-free images in uncorrected (G) and in motion-corrected images using the pre-analysis approach (H) and post-analysis approach (I). Every cell in A, B, C, G, H and I contains results from all 10 patients. Every cell in D, E and F contains results from the one patient with the highest maximum deviation
Figure 3
Figure 3
Median residual motion calculated as the difference between the actual simulated motion and the motion corrected by the two motion correction approaches. Positive values indicate under-correction, while negative values indicate over-correction. The values represent the magnitude of the residual motion in millimeters
Figure 4
Figure 4
Polar plots from a patient scan comparing images pre-correction (A), post-correction (B), and original motion-free (C). The simulated linear slide motion (20 mm, 1 minute post-injection) increased MBF in the anterior part and reduced MBF in the inferior part with a large apical inferolateral defect (A). After motion correction, the artifact disappeared (B), aligning with the original motion-free image (C)
Figure 5
Figure 5
PET images representing a patient from the clinical cohort pre- (A, B) and post-motion (C, D) correction. In the polar plot in A, a motion artifact has caused a false positive defect (20.1 %) in the RCA territory. In the post-motion correction polar plot in C, the motion artifact is completely reduced (defect 0.0 %). Splash images of short axis, horizontal long axis, and vertical long axis in B demonstrate the effect of motion artifacts on the visual interpretation of the images. The inferior wall appears hypoperfused compared to the rest of the myocardium. The post-motion correction splash images in D, demonstrate a uniform tracer uptake, thereby eliminating any suspicions of defects in the inferior wall. MBF, myocardial blood flow (mL⋅min−1⋅g−1); PTF, perfusable tissue fraction (mL⋅mL−1)
Figure 6
Figure 6
Scatter plots and Bland–Altman plots of regional MBFstress as well as global MBFstress for precorrection vs motion-corrected images in the clinical patient cohort. Dashed lines in the scatter plots indicate the threshold of normal values (MBF ≥ 2.3 mL⋅min−1⋅g−1). Red lines represent lines of identity. Dashed lines in the Bland–Altman plots represent limits of agreement

References

    1. Knaapen P, Lubberink M. Cardiac positron emission tomography: Myocardial perfusion and metabolism in clinical practice. Clin Res Cardiol. 2008;97:791–796. doi: 10.1007/s00392-008-0662-9. - DOI - PubMed
    1. Danad I, Uusitalo V, Kero T, Saraste A, Raijmakers PG, Lammertsma AA, et al. Quantitative assessment of myocardial perfusion in the detection of significant coronary artery disease: Cutoff values and diagnostic accuracy of quantitative [(15)O]H2O PET imaging. J Am Coll Cardiol. 2014;64:1464–1475. doi: 10.1016/j.jacc.2014.05.069. - DOI - PubMed
    1. Harms HJ, Knaapen P, de Haan S, Halbmeijer R, Lammertsma AA, Lubberink M. Automatic generation of absolute myocardial blood flow images using [15O]H2O and a clinical PET/CT scanner. Eur J Nucl Med Mol Imaging. 2011;38:930–939. doi: 10.1007/s00259-011-1730-3. - DOI - PMC - PubMed
    1. Murthy VL, Bateman TM, Beanlands RS, Berman DS, Borges-Neto S, Chareonthaitawee P, et al. Clinical quantification of myocardial blood flow using PET: Joint position paper of the SNMMI Cardiovascular Council and the ASNC. J Nucl Cardiol. 2018;25:269–297. doi: 10.1007/s12350-017-1110-x. - DOI - PubMed
    1. Naum A, Laaksonen MS, Tuunanen H, Oikonen V, Teras M, Kemppainen J, et al. Motion detection and correction for dynamic (15)O-water myocardial perfusion PET studies. Eur J Nucl Med Mol Imaging. 2005;32:1378–1383. doi: 10.1007/s00259-005-1846-4. - DOI - PubMed

Publication types

LinkOut - more resources