. 2023 Dec;55(S 01):E1035-E1038.
doi: 10.1055/a-2155-5608. Epub 2023 Sep 15.

Advantages of the dual-channel multi-bending endoscope for ERCP in patients with Billroth II reconstruction

Affiliations

Advantages of the dual-channel multi-bending endoscope for ERCP in patients with Billroth II reconstruction

Haruka Toyonaga et al. Endoscopy. 2023 Dec.
No abstract available

PubMed Disclaimer

Conflict of interest statement

A. Katanuma has received honoraria as a lecture fee from Olympus Co., Tokyo, Japan. H. Toyonaga, T. Hayashi, M. Motoya, T. Kin, and K. Takahashi declare that they have no conflict of interest.

Figures

Fig. 1
Fig. 1
The M-scope (GIF-2TQ260M; Olympus Corp., Tokyo, Japan). a The scope has dual working channels and two bending sites (Scope tip diameter φ11.7 mm, channel diameter φ3.2 mm/φ3.2 mm). The two bending sites allow for a variety of scope positions: b first angle up and second angle neutral; c first angle up and second angle up; d first angle up and second angle down; e first angle fully up and second angle down. White arrow, first bending site; blue arrow, second bending site.
Fig. 2
Fig. 2
The endoscope can negotiate the acute angles encountered in patients with surgically altered anatomy. a Moving only the first angle upward cannot overcome the acute angle, and leads only to extension of the intestine. b,c By adjusting the second angle upwards, the acute angle was easily negotiated and the scope tip was guided more deeply. d In order to round the next bend, the first bending site was angled downwards to capture the lumen. e The second bending site was set to neutral, enabling deeper insertion of the scope tip.  f The short position of the scope allows stability. White arrow, first bending site; blue arrow, second bending site.
Fig. 3
Fig. 3
Endoscopic and fluoroscopic views. a, b In the short scope position, the endoscope and the papilla are tangential, and the biliary orifice is not visible. c, d During maneuvering of the first bending site, the endoscope tip fails to maintain a sufficient distance from the papilla and becomes embedded in the mucosa; the field of view cannot be secured. e, f By manipulating the second bending site in combination with the first, it is possible to obtain an appropriate distance while maintaining a frontal view of the papilla. Asterisk, papilla; white arrow, first bending site; blue arrow, second bending site.
Fig. 4
Fig. 4
Use of the scope during cannulation. a–c Using the two channels of the endoscope, forceps inserted through one channel can be used to hold and pull the papilla during biliary cannulation, and a catheter can be inserted through the other channel.
Fig. 5
Fig. 5
Illustration depicting a needle-knife sphincterotomy and endoscopic papillary large balloon dilation (EPLBD) using the multi-bending maneuver. a, b Using only the first bending position, the tip of the knife deviates from the submucosa with each incision. In addition, moving the needle tip with the push/pull of the scope may produce unpredictable results, particularly in patients with surgically altered anatomy. c By utilizing both the first and second bending sites, an appropriate angle can be created without moving the scope push/pull, thereby allowing for an efficient and safe incision. d, e Inflation of the EPLBD balloon creates a force that pulls the balloon into the bile duct (green arrow). To adjust to the proper position, the scope tip is moved away by pulling the scope and manipulating the angle of the first bend. However, it is difficult to exert sufficient force to pull the balloon, making fine adjustments difficult. f The second bending site can be used in conjunction with the first to fine-tune the distance of the scope tip without pulling the scope, thereby allowing the balloon position to be adjusted while maintaining a stable scope position. White circle, first bending site; blue circle, second bending site.

References

    1. Katanuma A, Yane K, Osanai M et al.Endoscopic retrograde cholangiopancreatography in patients with surgically altered anatomy using balloon-assisted enteroscope. Clin J Gastroenterol. 2014;7:283–289. - PubMed
    1. Yane K, Katanuma A, Maguchi H et al.Short-type single-balloon enteroscope-assisted ERCP in postsurgical altered anatomy: potential factors affecting procedural failure. Endoscopy. 2017;49:69–74. - PubMed
    1. Okabe Y, Ishida Y, Kuraoka K et al.Endoscopic bile duct and/or pancreatic duct cannulation technique for patients with surgically altered gastrointestinal anatomy. Dig Endosc. 2014;26 02:122–126. - PubMed
    1. Koo H C, Moon J H, Choi H J et al.The utility of a multibending endoscope for selective cannulation during ERCP in patients with a Billroth II gastrectomy (with video) Gastrointest Endosc. 2009;69:931–934. - PubMed

MeSH terms