. 2023 Sep 18;199(14):1616-1619.
doi: 10.1093/rpd/ncad127.

Alanine/EPR dosimetry for mailed intercomparison at ocular proton therapy facilities-preliminary results for three centres for irradaitions at CCB IFJ PAN eyeline

Affiliations

Alanine/EPR dosimetry for mailed intercomparison at ocular proton therapy facilities-preliminary results for three centres for irradaitions at CCB IFJ PAN eyeline

Barbara Michalec et al. Radiat Prot Dosimetry. .

Abstract

Quality control of therapeutic photon beams in the form of postal dose audits based on passive dosemeters is widely used in photon radiotherapy. On the other hand, no standardised dosimetry audit programme for proton centres has been established in Europe so far. We evaluated alanine/EPR dosimetry systems developed at the Istituto Superiore di Sanità (Italy), the Hasselt Universiteit (Belgium) and the Henryk Niewodniczanski Institute of Nuclear Physics Polish Academy of Sciences (Poland) for their applicability as a potential tool for routine mailed dose audits of passively scattered therapeutic proton beams. The evaluation was carried out in the form of an intercomparison. Dosemeters were irradiated in the 70 MeV proton beam at ocular proton therapy facility in the Cyclotron Centre Bronowice at the Henryk Niewodniczanski Institute of Nuclear Physics Polish Academy of Sciences in Krakow. A very good agreement was found between the dose measured by three laboratories and the delivered dose determined with an ionisation chamber. This, together with the inherent properties of alanine, such as non-destructive readout, tissue equivalence, weak energy dependence, dose rate independence and insignificant fading, makes alanine a good candidate for a dosemeter used in postal auditing in proton ocular radiotherapy.

PubMed Disclaimer

Conflict of interest statement

The authors declare no conflict of interest.

Figures

Figure 1
Figure 1
The scheme of the PMMA phantom used in the intercomparison for alanine irradiation. A - PMMA phantom; cover + base, B - cover bottom view, C - cover side view.
Figure 2
Figure 2
Intercomparison irradiation of the phantom with alanine dosemeters.
Figure 3
Figure 3
The mean alanine dose reported in the intercomparison by ISS, HU and IFJ PAN.

References

    1. International Atomic Energy Agency . Comprehensive Audits of Radiotherapy Practices: A Tool for Quality Improvement (Vienna: IAEA, Non-serial Publications; ) (2007).
    1. Ferreira, I. H., Dutreix, A., Bridier, A., Chavaudra, J. and Svensson, H. The ESTRO-QUALity assurance network (EQUAL). Radiother. Oncol. 55(3), 273–284 (2000). - PubMed
    1. Rozporządzenie Ministra Zdrowia z dnia 18 lutego 2011 r. w sprawie warunków bezpiecznego stosowania promieniowania jonizującego dla wszystkich rodzajów ekspozycji medycznej (t.j. Dz. U. z 2017 r. poz. 884).
    1. Tian, X., Liu, K., Hou, Y., Cheng, J. and Zhang, J. The evolution of proton beam therapy: current and future status. Mol. Clin. Oncol. 8(1), 15–21. - PMC - PubMed
    1. Carlino, A., Gouldstone, C., Kragl, G., Traneus, E., Marrale, M., Vatnitsky, S., Stock, M. and Palmans, H. End-to-end tests using alanine dosimetry in scanned proton beams. Phys. Med. Biol. 63(5), 055001 (2018). - PubMed