. 2023 Sep 19;12(18):6054.
doi: 10.3390/jcm12186054.

Accuracy-Based Glomerular Filtration Rate Assessment by Plasma Iohexol Clearance in Kidney Transplant Donors

Affiliations

Accuracy-Based Glomerular Filtration Rate Assessment by Plasma Iohexol Clearance in Kidney Transplant Donors

Zhicheng Jin et al. J Clin Med. .

Abstract

Background: An accurate measurement of the glomerular filtration rate (GFR) is essential for detecting renal insufficiency in living kidney donors. Iohexol is a "near-ideal" exogenous filtration marker for GFR measurements that has attracted increasing interest in clinical practice because it is non-toxic, non-radioactive, readily available, and easy to measure. In this study, we aimed to set up a laboratory test to conveniently assess the plasma clearance of iohexol in living kidney donors.

Methods: A workflow was established in the institution's infusion clinic to administer iohexol and to collect three timed blood samples from renal transplant donors. Iohexol was thereafter measured by liquid chromatography-tandem mass spectrometry (LC-MS/MS). The serum proteins were precipitated and the supernatant containing iohexol was diluted prior to the LC-MS/MS analysis. The LC-MS/MS method was developed on a Thermo Vanquish UHPLC coupled with a TSQ Endura triple quadruple mass spectrometer with a total run time of 2.5 min. The analytical performance of the method was assessed.

Results: The LC-MS/MS method demonstrated a good analytical performance. To calculate the iohexol clearance rate and the GFR, automated data integration and a result calculation were accomplished by using a custom Python script. Automated result reporting was achieved using a laboratory informatics system (LIS) vendor's direct media interface.

Conclusions: We developed and implemented a laboratory test to assess the plasma clearance of iohexol. A workflow was established in the hospital to reliably measure the GFR in living kidney donors, with a potential to be further expanded into other areas where an accurate GFR measurement is needed.

Keywords: LC-MS/MS; glomerular filtration rate; plasma iohexol clearance.

PubMed Disclaimer

Conflict of interest statement

The authors declare no conflict of interest.

Figures

Figure 1
Figure 1
Extracted ion chromatograms of iohexol and d5-iohexol.
Figure 2
Figure 2
Determination of lower limit of quantitation (LOQ).
Figure 3
Figure 3
Method comparison study analysis using Deming regression (A) and Bland–Altman plots (B).

References

    1. Levey A.S., Stevens L.A., Schmid C.H., Zhang Y.L., Castro A.F., 3rd, Feldman H.I., Kusek J.W., Eggers P., Van Lente F., Greene T., et al. A New Equation to Estimate Glomerular Filtration Rate. Ann. Intern. Med. 2009;150:604–612. doi: 10.7326/0003-4819-150-9-200905050-00006. - DOI - PMC - PubMed
    1. Porrini E., Ruggenenti P., Luis-Lima S., Carrara F., Jiménez A., de Vries A.P.J., Torres A., Gaspari F., Remuzzi G. Estimated GFR: Time for a critical appraisal. Nat. Rev. Nephrol. 2019;15:177–190. doi: 10.1038/s41581-018-0080-9. - DOI - PubMed
    1. Gaspari F., Perico N., Ruggenenti P., Mosconi L., Amuchastegui C.S., Guerini E., Daina E., Remuzzi G. Plasma Clearance of Nonradioactive Iohexol as a Measure of Glomerular Filtration Rate. J. Am. Soc Nephrol. 1995;6:257–263. doi: 10.1681/ASN.V62257. - DOI - PubMed
    1. Delanaye P., Melsom T., Ebert N., Bäck S.-E., Mariat C., Cavalier E., Björk J., Christensson A., Nyman U., Porrini E., et al. Iohexol plasma clearance for measuring glomerular filtration rate in clinical practice and research: A review. Part 2: Why to measure glomerular filtration rate with iohexol? Clin. Kidney J. 2016;9:700–704. doi: 10.1093/ckj/sfw071. - DOI - PMC - PubMed
    1. Luis-Lima S., Porrini E. An Overview of Errors and Flaws of Estimated GFR versus True GFR in Patients with Diabetes Mellitus. Nephron. 2017;136:287–291. doi: 10.1159/000453531. - DOI - PubMed

LinkOut - more resources