Underestimation of Occupational Radiation Exposure During Endovascular Abdominal Aortic Aneurysm Repair
- PMID: 37819216
- DOI: 10.1097/HP.0000000000001744
Underestimation of Occupational Radiation Exposure During Endovascular Abdominal Aortic Aneurysm Repair
Abstract
During interventional procedures of endovascular abdominal aortic aneurysm repair (EVAR), the dosimeter was conventionally placed on chest facing toward the surgical table, instead of the main source of scatter radiation. Purpose of this study is to evaluate the underestimation of occupational radiation exposure. Phantom experiments were performed in a hybrid operating room equipped with an interventional angiography system. Electric personal dosimeters were placed at the level of eyes, chest, abdomen, and gonad of three positions, representing the principal operator (PO), assistant operator (AO), and sterile nurse (SN). Personal dose equivalent was measured with two different orientations of radiation detection, facing the table and facing the phantom, respectively. In addition to fluoroscopy, the dose produced by digital subtraction angiography was also measured to estimate the radiation exposure of routine EVAR. In this study, staff doses of 26 EVAR cases were also collected in our hospital to correlate the estimated dose. Our results show that the facing-phantom dose normalized by dose area product of patient is significantly higher than the facing-table dose when the latter is regularly seen in clinical practice. This underestimation could be even worse at a more distant position (e.g., AO and SN) as the incident angle of scatter radiation is larger. Besides, the estimated dose is highly correlated with the on-site measured dose (R 2 ~ 0.8) at chest and gonad of the PO.
Copyright © 2023 Health Physics Society.
References
-
- Boyle JR, Mao J, Beck AW, Venermo M, Sedrakyan A, Behrendt CA, Szeberin Z, Eldrup N, Schermerhorn M, Beiles B, Thomson I, Cassar K, Altreuther M, Debus S, Johal AS, Waton S, Scali ST, Cromwell DA, Mani K. Editor's choice—variation in intact abdominal aortic aneurysm repair outcomes by country: analysis of international consortium of vascular registries 2010–2016. Eur J Vasc Endovasc Surg 62:16–24; 2021.
-
- Chiriotti S, Ginjaume M, Vano E, Sanchez R, Fernandez JM, Duch MA, Sempau J. Performance of several active personal dosemeters in interventional radiology and cardiology. Radiat Meas 46:1266–1270; 2011.
-
- Clairand I, Bordy JM, Daures J, Debroas J, Denoziere M, Donadille L, Ginjaume M, Itie C, Koukorava C, Krim S, Lebacq AL, Martin P, Struelens L, Sans-Merce M, Tosic M, Vanhavere F. Active personal dosemeters in interventional radiology: tests in laboratory conditions and in hospitals. Radiat Protect Dosim 144:453–458; 2011.
-
- de Ruiter QMB, Jansen MM, Moll FL, Hazenberg C, Kahya NN, van Herwaarden JA. Procedure and step-based analysis of the occupational radiation dose during endovascular aneurysm repair in the hybrid operating room. J Vasc Surg 67:1881–1890; 2018.
-
- Owens DK, Davidson KW, Krist AH, Barry MJ, Cabana M, Caughey AB, Doubeni CA, Epling JW Jr., Kubik M, Landefeld CS, Mangione CM, Pbert L, Silverstein M, Simon MA, Tseng CW, Wong JBForce USPST. Screening for abdominal aortic aneurysm: US preventive services task force recommendation statement. JAMA 322:2211–2218; 2019.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical