Radiofrequency lesion formation prediction with contact force versus local impedance
Abstract
Purpose of review: Safe and effective radiofrequency (RF) myocardial ablation requires real-time monitoring of lesion formation. Here, we review conventional and novel approaches proposed to guide titration of RF energy application.
Recent findings: Conventional monitoring modalities, such as ablation electrode temperature, generator impedance, and tissue electrophysiological properties have been of limited value in predicting efficacy and safety of ablation. Therefore, several input-driven indices have been proposed to improve the quality and durability of RF ablation lesion while maintaining safety. These metrics predominantly incorporate RF power output, duration of RF application, and firmness and stability of electrode-tissue contact. More recently, novel catheters have enabled measuring local impedance at the catheter-tissue interface, which has been found valuable for real-time monitoring of RF lesion formation.
Summary: It is likely that using the combination of multiple metrics would be required to improve the quality and safety of RF lesions, but further investigation is still required.
Copyright © 2023 Wolters Kluwer Health, Inc. All rights reserved.
References
-
- Issa ZF, Miller JM, Zipes DP. Issa ZF, Miller JM, Zipes DP. Ablation energy sources. Clinical arrhythmology and electrophysiology: a companion to Braunwald's heart disease 3rd ed.Philadelphia, PA, USA: Elsevier; 2019. 206–237.
-
- Habibi M, Berger RD, Calkins H. Radiofrequency ablation: technological trends, challenges, and opportunities. Europace 2021; 23:511–519.
-
- Nguyen DT, Zipse M, Borne RT, et al. Use of tissue electric and ultrasound characteristics to predict and prevent steam-generated cavitation during high-power radiofrequency ablation. JACC Clin Electrophysiol 2018; 4:491–500.
-
- Mansour M, Calkins H, Osorio J, et al. Persistent atrial fibrillation ablation with contact force-sensing catheter: the prospective multicenter PRECEPT Trial. JACC Clin Electrophysiol 2020; 6:958–969.
-
- Conti S, Weerasooriya R, Novak P, et al. Contact force sensing for ablation of persistent atrial fibrillation: a randomized, multicenter trial. Heart Rhythm 2018; 15:201–208.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources