EyeArt artificial intelligence analysis of diabetic retinopathy in retinal screening events
- PMID: 37847478
- DOI: 10.1007/s10792-023-02887-9
EyeArt artificial intelligence analysis of diabetic retinopathy in retinal screening events
Abstract
Purpose: Early detection and treatment of diabetic retinopathy (DR) are critical for decreasing the risk of vision loss and preventing blindness. Community vision screenings may play an important role, especially in communities at higher risk for diabetes. To address the need for increased DR detection and referrals, we evaluated the use of artificial intelligence (AI) for screening DR.
Methods: Patient images of 124 eyes were obtained using a 45° Canon Non-Mydriatic CR-2 Plus AF retinal camera in the Department of Endocrinology Clinic (Newark, NJ) and in a community screening event (Newark, NJ). Images were initially classified by an onsite grader and uploaded for analysis by EyeArt, a cloud-based AI software developed by Eyenuk (California, USA). The images were also graded by an off-site retina specialist. Using Fleiss kappa analysis, a correlation was investigated between the three grading systems, the AI, onsite grader, and a US board-certified retina specialist, for a diagnosis of DR and referral pattern.
Results: The EyeArt results, onsite grader, and the retina specialist had a 79% overall agreement on the diagnosis of DR: 86 eyes with full agreement, 37 eyes with agreement between two graders, 1 eye with full disagreement. The kappa value for concordance on a diagnosis was 0.69 (95% CI 0.61-0.77), indicating substantial agreement. Referral patterns by EyeArt, the onsite grader, and the ophthalmologist had an 85% overall agreement: 96 eyes with full agreement, 28 eyes with disagreement. The kappa value for concordance on "whether to refer" was 0.70 (95% CI 0.60-0.80), indicating substantial agreement. Using the board-certified retina specialist as the gold standard, EyeArt had an 81% accuracy (101/124 eyes) for diagnosis and 83% accuracy (103/124 eyes) in referrals. For referrals, the sensitivity of EyeArt was 74%, specificity was 87%, positive predictive value was 72%, and negative predictive value was 88%.
Conclusions: This retrospective cross-sectional analysis offers insights into use of AI in diabetic screenings and the significant role it will play in automated detection of DR. The EyeArt readings were beneficial with some limitations in a community screening environment. These limitations included a decreased accuracy in the presence of cataracts and the functional cost of EyeArt uploads in a community setting.
Keywords: Artificial intelligence; Community vision outreach; Diabetic retinopathy.
© 2023. The Author(s), under exclusive licence to Springer Nature B.V.
Similar articles
-
Artificial Intelligence Detection of Diabetic Retinopathy: Subgroup Comparison of the EyeArt System with Ophthalmologists' Dilated Examinations.Ophthalmol Sci. 2022 Sep 30;3(1):100228. doi: 10.1016/j.xops.2022.100228. eCollection 2023 Mar. Ophthalmol Sci. 2022. PMID: 36345378 Free PMC article.
-
The Value of Automated Diabetic Retinopathy Screening with the EyeArt System: A Study of More Than 100,000 Consecutive Encounters from People with Diabetes.Diabetes Technol Ther. 2019 Nov;21(11):635-643. doi: 10.1089/dia.2019.0164. Epub 2019 Aug 7. Diabetes Technol Ther. 2019. PMID: 31335200 Free PMC article.
-
Telemedical Diabetic Retinopathy Screening in a Primary Care Setting: Quality of Retinal Photographs and Accuracy of Automated Image Analysis.Ophthalmic Epidemiol. 2022 Jun;29(3):286-295. doi: 10.1080/09286586.2021.1939886. Epub 2021 Jun 20. Ophthalmic Epidemiol. 2022. PMID: 34151725
-
Review of retinal cameras for global coverage of diabetic retinopathy screening.Eye (Lond). 2021 Jan;35(1):162-172. doi: 10.1038/s41433-020-01262-7. Epub 2020 Nov 9. Eye (Lond). 2021. PMID: 33168977 Free PMC article. Review.
-
Various models for diabetic retinopathy screening that can be applied to India.Indian J Ophthalmol. 2021 Nov;69(11):2951-2958. doi: 10.4103/ijo.IJO_1145_21. Indian J Ophthalmol. 2021. PMID: 34708729 Free PMC article. Review.
Cited by
-
Distribution of Microaneurysms and Hemorrhages in Accordance with the Grading of Diabetic Retinopathy in Type Diabetes Patients.Diagnostics (Basel). 2024 Jul 17;14(14):1547. doi: 10.3390/diagnostics14141547. Diagnostics (Basel). 2024. PMID: 39061684 Free PMC article.
-
The Use of Artificial Intelligence for Estimating Anterior Chamber Depth from Slit-Lamp Images Developed Using Anterior-Segment Optical Coherence Tomography.Bioengineering (Basel). 2024 Oct 9;11(10):1005. doi: 10.3390/bioengineering11101005. Bioengineering (Basel). 2024. PMID: 39451381 Free PMC article.
-
Comparison of Validity and Reliability of Manual Consensus Grading vs. Automated AI Grading for Diabetic Retinopathy Screening in Oslo, Norway: A Cross-Sectional Pilot Study.J Clin Med. 2025 Jul 7;14(13):4810. doi: 10.3390/jcm14134810. J Clin Med. 2025. PMID: 40649184 Free PMC article.
-
The application of artificial intelligence in diabetic retinopathy: progress and prospects.Front Cell Dev Biol. 2024 Oct 25;12:1473176. doi: 10.3389/fcell.2024.1473176. eCollection 2024. Front Cell Dev Biol. 2024. PMID: 39524224 Free PMC article. Review.
References
-
- Prevention CFDCa (2022) National diabetes statistics report. National Diabetes Statistics Report Website
-
- Control CFD (2020) Prevalence of prediabetes among adults. Diabetes. CDC
-
- National Diabetes Statistics Report (2020) Centers for disease control and prevention
-
- Gruss SM, Nhim K, Gregg E, Bell M, Luman E, Albright A (2019) Public health approaches to type 2 diabetes prevention: the US National Diabetes Prevention Program and beyond. Curr Diabetes Rep 19:1–11 - DOI
-
- Lawrenson JG, Graham-Rowe E, Lorencatto F, Burr J, Bunce C, Francis JJ et al (2018) Interventions to increase attendance for diabetic retinopathy screening. Cochrane Database Syst Rev 1:CD012054 - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Medical