The field dependence of NMR imaging. I. Laboratory assessment of signal-to-noise ratio and power deposition
- PMID: 3784889
- DOI: 10.1002/mrm.1910030508
The field dependence of NMR imaging. I. Laboratory assessment of signal-to-noise ratio and power deposition
Abstract
A method is proposed for measuring on the bench the NMR signal-to-noise ratio of rf probes, (over the range 1-100 MHz) and also the power deposited in patients during the imaging experiment. The technique is based on the principle of reciprocity, in that a direct relationship exists between the magnetic field generated (upon transmission) by a matched probe coil and the signal-to-noise ratio delivered by the same coil when used as a receiver. The construction and use of a calibrated sense coil for measuring the field is described, and the precautions and theory necessary for accurate measurement and understanding are outlined. Finally, the method is verified by comparison with a direct spectral measure of sensitivity obtained from a small doped water sample placed in NMR imaging equipment.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
