Development of a motorized system for the characterization of the RQR reference x-ray beams at the KFSHRC's SSDL using an extrapolation chamber
- PMID: 37866227
- DOI: 10.1016/j.apradiso.2023.111064
Development of a motorized system for the characterization of the RQR reference x-ray beams at the KFSHRC's SSDL using an extrapolation chamber
Abstract
An extrapolation chamber of type Böhm 23392, used for the dosimetry of RQR beam qualities at the Secondary Standard Dosimetry Laboratory of King Faisal Specialist Hospital, is presented. A computer-based motorized system consisting of a stepping motor coupled to the chamber's built-in micrometer screw was designed to expedite the measurements process, giving a linear relationship between the number of motor steps and the chamber depth with a Coefficient of Determination (COD) equal to 0.9990. The extrapolation chamber along with the motorized system was used to determine the extrapolation curves. The slope of the extrapolation curves, measured for the eight RQR beam qualities with a relative uncertainty between 0.17% and 0.58%, varies linearly with the beam quality expressed in terms of HVL, with a COD between 0.9995 and 1.0000.
Copyright © 2023 Elsevier Ltd. All rights reserved.
Conflict of interest statement
Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
LinkOut - more resources