Review
. 2023 Nov;47(11):648-657.
doi: 10.1016/j.medine.2023.10.009. Epub 2023 Oct 20.

Reverse triggering during mechanical ventilation: Diagnosis and clinical implications

Affiliations
Review

Reverse triggering during mechanical ventilation: Diagnosis and clinical implications

Juan M Núñez Silveira et al. Med Intensiva (Engl Ed). 2023 Nov.

Abstract

This review addresses the phenomenon of "reverse triggering", an asynchrony that occurs in deeply sedated patients or patients in transition from deep to light sedation. Reverse triggering has been reported to occur in 30-90% of all ventilated patients. The underlying pathophysiological mechanisms remain unclear, but "entrainment" is proposed as one of them. Detecting this asynchrony is crucial, and methods such as visual inspection, esophageal pressure, diaphragmatic ultrasound and automated methods have been used. Reverse triggering may have effects on lung and diaphragm function, probably mediated by the level of breathing effort and eccentric activation of the diaphragm. The optimal management of reverse triggering has not been established, but may include the adjustment of ventilatory parameters as well as of sedation level, and in extreme cases, neuromuscular block. It is important to understand the significance of this condition and its detection, but also to conduct dedicated research to improve its clinical management and potential effects in critically ill patients.

Keywords: Asincronía paciente-ventilador; Mechanical ventilation; Patient-ventilator asynchrony; Reverse triggering; Trigger reverso; Ventilación mecánica.