Paroxysmal atrial fibrillation ablation with a novel temperature-controlled CF-sensing catheter: Q-FFICIENCY clinical and healthcare utilization benefits
- PMID: 37870157
- DOI: 10.1111/jce.16093
Paroxysmal atrial fibrillation ablation with a novel temperature-controlled CF-sensing catheter: Q-FFICIENCY clinical and healthcare utilization benefits
Abstract
Introduction: The prospective, nonrandomized, multicenter Q-FFICIENCY study demonstrated the safety and 12-month efficacy of paroxysmal atrial fibrillation (AF) ablation with the novel QDOT MICRO temperature-controlled, contact force-sensing, radiofrequency (RF) catheter. Participants underwent pulmonary vein isolation with very high-power short-duration (vHPSD) mode (90 W, ≤4 s) alone or combined with conventional-power temperature-controlled (CPTC) mode (25-50 W). This study aimed to assess quality-of-life (QOL) and healthcare utilization (HCU) benefits experienced by Q-FFICIENCY study participants.
Methods: Besides evaluating procedural efficiency, QOL and HCU were assessed through 12 months postablation via Atrial Fibrillation Effect on Quality-of-Life Tool (AFEQT) score, antiarrhythmic drug (AAD) use, and incidence of cardioversion and cardiovascular hospitalization.
Results: Of 191 participants enrolled, 166 were ablated with the new catheter. Compared to baseline, statistically significant, clinically meaningful improvements in composite and subcategories of AFEQT scores were observed at 3 months and sustained through 12 months (12-month increase, 29.3-44.2 points). Class I/III AAD use decreased from 97.6% (162/166) at baseline to 19.6% (31/158) during Months 6-12, representing a significant 79.9% reduction. The cardioversion rate significantly declined by 93.9% from 31.3% (12 months preablation) to 1.9% (evaluation period). One-year Kaplan-Meier estimates of freedom from all-cause and cardiovascular hospitalization were 80.9% (95% confidence interval [CI], 74.8%-86.9%) and 88.8% (95% CI, 84.0%-93.7%), respectively.
Conclusions: Paroxysmal AF ablation with the novel temperature-controlled RF catheter in vHPSD mode, alone or with CPTC mode, led to clinically meaningful improvement in QOL and significant reduction in AAD use, cardioversion, and cardiovascular hospitalization.
Keywords: QDOT MICRO; QMODE+; healthcare utilization; quality of life; temperature-controlled ablation; very high-power short-duration.
© 2023 The Authors. Journal of Cardiovascular Electrophysiology published by Wiley Periodicals LLC.
References
REFERENCES
-
- Calkins H, Hindricks G, Cappato R, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Heart Rhythm. 2017;14:e275-e444.
-
- Ladwig KH, Goette A, Atasoy S, Johar H. Psychological aspects of atrial fibrillation: a systematic narrative review: impact on incidence, cognition, prognosis, and symptom perception. Curr Cardiol Rep. 2020;22:137.
-
- Tsao CW, Aday AW, Almarzooq ZI, et al. Heart disease and stroke statistics-2022 update: a report from the American Heart Association. Circulation. 2022;145:e153-e639.
-
- Mark DB, Anstrom KJ, Sheng S, et al. CABANA Investigator Effect of catheter ablation vs medical therapy on quality of life among patients with atrial fibrillation: the CABANA randomized clinical trial. JAMA. 2019;321:1275-1285.
-
- Hindricks G, Potpara T, Dagres N, et al. 2020 ESC guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): the task force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J. 2021;42:373-498.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Medical