Innovations in HbA1c analysis: finding the balance between speed and accuracy. An investigation of a potential new Secondary Reference Measurement Procedure for the IFCC
- PMID: 37870921
- DOI: 10.1515/cclm-2023-1070
Innovations in HbA1c analysis: finding the balance between speed and accuracy. An investigation of a potential new Secondary Reference Measurement Procedure for the IFCC
Abstract
Objectives: The escalating prevalence of diabetes worldwide has resulted in a dramatic increase in the number of people who need testing, which in turn necessitates faster HbA1c measurement. The Tosoh GR01 addresses the need for fast turnaround times of whilst offering pragmatic steps to maintain result accuracy in a single instrument by offering two distinct operating modes: Short Mode (SM) and Long Mode (LM). The aim of this study was to evaluate all relevant aspects of the performance of the Tosoh GR01 with a view to accepting the instrument as a future Secondary Reference Measurement Procedure (SRMP) for the IFCC.
Methods: Certified Clinical & Laboratory Standards Institute (CLSI) Evaluation Protocols (EP) were used to evaluate precision (EP-5), accuracy (EP-9), linearity (EP-6), carry-over (EP-10) and the effect of hemoglobin variants and other potential interferences.
Results: Both modes demonstrated CVs <0.6 % in SI units and <0.4 % in NGSP units at 46 mmol/mol (6.4 %) and 75 mmol/mol (9.0 %) and passed both National Glycohemoglobin Standardization Program (NGSP) and International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) certification procedures when compared with 7 IFCC Certified Secondary Reference Measurement Procedures (SRMP). Sigma for both modes was >6 when using the results of EP-5 and EP-9 at an HbA1c concentration of 50 mmol/mol (6.7 %). Neither mode showed any interference with common Hb-variants except for HbAE when HbA1c was >65 mmol/mol. In the SM HbAS, HbAD and HbAC were recognized but no result was reported.
Conclusions: There is a good balance between speed and accuracy for determining HbA1c with the Tosoh GR01 in both analytical modes and the device is suitable for use as an IFCC SRMP.
Keywords: HbA1c; Tosoh GR01; diabetes; method evaluation.
© 2023 Walter de Gruyter GmbH, Berlin/Boston.
References
-
- DCCT Research Group . The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med 1993;329:977–86. https://doi.org/10.1056/NEJM199309303291401 . - DOI
-
- United Kingdom Prospective Diabetes Study (UKPDS) . Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes. Lancet 1998;352:837–53.
-
- Weykamp, C, John, G, Gillery, P, English, E, Ji, L, Lenters-Westra, E, et al.. Investigation of 2 models to set and evaluate quality targets for HbA1c: biological variation and sigma-metrics. Clin Chem 2015;61:752–9. https://doi.org/10.1373/clinchem.2014.235333 . - DOI
-
- International Diabetes Federation diabetes atlas prevalence diabetes [Online]. https://diabetesatlas.org/ [Accessed 21 Sep 2023].
-
- Sacks, DB, Arnold, M, Bakris, GL, Bruns, DE, Horvath, AR, Lernmark, Ä, et al.. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem 2023;69:808–68. https://doi.org/10.1093/clinchem/hvad080 . - DOI
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous