. 2024 Jan;43(1):153-160.
doi: 10.1002/nau.25317. Epub 2023 Oct 27.

Development and construction of an open-platform, open-source low-cost portable uroflowmeter: The OpenFlow device

Affiliations

Development and construction of an open-platform, open-source low-cost portable uroflowmeter: The OpenFlow device

Germano Adelino Gallo et al. Neurourol Urodyn. 2024 Jan.

Abstract

Background: Uroflowmetry is useful to screen for and manage many voiding disorders. Home-based uroflowmetry might better represent the patient's true voiding pattern and be more widely adopted if an accurate low-cost portable device was available.

Objective: Development and initial evaluation of an open-platform, open-source low-cost portable uroflowmeter.

Materials and methods: We designed and built an uroflowmeter comprising of a load cell and digital memory card unit connected to a programmable microcontroller board mounted upon a 3D printed frame. It generated date-stamped tables which were processed and plotted. Twenty urologists were recruited to assess the device. Each participant received the equipment that was returned, along with a bladder diary, after at least 24 consecutive hours of homemade uroflowmetry recording. Additionally, were assessed with the International Prostatic Symptom Score (I-PSS) and Peeling diagram, whereas the device's ease of use, robustness, and portability were evaluated with a Likert-type questionnaire. Two experienced urodynamicists independently evaluated the tracings' quality rated with a 3° ordinal scale: (1) Interpretable without artifacts; (2) Interpretable with artifacts; (3) Uninterpretable.

Results: Participants' median age was 36.6 years old, none having an I-PSS > 5 or Peeling > 2. Overall 138 voidings were recorded (77 daytime, 61 nightly episodes). The device's ease of use, robustness, and portability obtained maximum score in 80% of evaluations. Most (98%) of the tracings were considered interpretable. Limitations included its small study population and short monitoring times.

Conclusion: The construction of a cheap (<50 USD), accurate user-friendly portable uroflowmeter proved feasible, which could facilitate access to portable uroflowmetry.

Keywords: home-based uroflowmetry; open-source software; portable uroflowmeter; uroflowmetry.

PubMed Disclaimer

References

REFERENCES

    1. Veeratterapillay R, Pickard R, Harding C. The role of uroflowmetry in the assessment and management of men with lower urinary tract symptoms - revisiting the evidence. J Clin Urol. 2014;7(3):154-158. doi:10.1177/2051415813498874
    1. Erickson BA, Breyer BN, McAninch JW. The use of uroflowmetry to diagnose recurrent stricture after urethral reconstructive surgery. J Urol. 2010;184(4):1386-1390. doi:10.1016/j.juro.2010.06.015
    1. Panicker JN. Neurogenic bladder: epidemiology, diagnosis, and management. Semin Neurol. 2020;40(5):569-579. doi:10.1055/s-0040-1713876
    1. Schäfer W, Abrams P, Liao L, et al. Good urodynamic practices: uroflowmetry, filling cystometry, and pressure-flow studies. Neurourol Urodyn. 2002;21(3):261-274. doi:10.1002/nau.10066
    1. Bray A, Griffiths C, Drinnan M, Pickard R. Methods and value of home uroflowmetry in the assessment of men with lower urinary tract symptoms: a literature review. Neurourol Urodyn. 2012;31(1):7-12. doi:10.1002/nau.21197

LinkOut - more resources