Effect of Implant Macro-Design and Magnetodynamic Surgical Preparation on Primary Implant Stability: An In Vitro Investigation
- PMID: 37886912
- PMCID: PMC10605303
- DOI: 10.3390/dj11100227
Effect of Implant Macro-Design and Magnetodynamic Surgical Preparation on Primary Implant Stability: An In Vitro Investigation
Abstract
Background: Macro-geometry and surgical implant site preparation are two of the main factors influencing implant stability and potentially determining loading protocol. The purpose of this study was to assess the initial stability of various implant macro-designs using both magnetodynamic and traditional osteotomy techniques in low-density bone. The parameters examined included peak insertion torque (PIT), implant stability quotient (ISQ), and peak removal torque (PRT).
Methods: Four groups of 34 implants each were identified in accordance with the surgery and implant shape: T5 group (Five implant and osteotomy using drills); M5 group (Five implant and magnetodynamic osteotomy using Magnetic Mallet); TT group (TiSmart implant and osteotomy with drills); and MT group (TiSmart implant and magnetodynamic osteotomy). Every implant was placed into a low-density bone animal model and scanned using CBCT. The PIT and PRT were digitally measured in Newton-centimeters (Ncm) using a torque gauge device. The ISQ was analyzed by conducting resonance frequency analysis.
Results: The PIT values were 25.04 ± 4.4 Ncm for T5, 30.62 ± 3.81 Ncm for M5, 30 ± 3.74 Ncm for TT, and 32.05 ± 3.55 Ncm for MT. The average ISQ values were 68.11 ± 3.86 for T5, 71.41 ± 3.69 for M5, 70.88 ± 3.08 for TT, and 73 ± 3.5 for MT. The PRT values were 16.47 ± 4.56 Ncm for T5, 26.02 ± 4.03 Ncm for M5, 23.91 ± 3.28 Ncm for TT, and 26.93 ± 3.96 Ncm for MT. Based on our data analysis using a t-test with α = 0.05, significant differences in PIT were observed between TT and T5 (p < 0.0001), M5 and T5 (p < 0.0001), and MT and TT (p = 0.02). Significant differences in the ISQ were found between TT and T5 (p = 0.001), M5 and T5 (p < 0.001), and MT and TT (p = 0.01). The PRT also exhibited significant differences between TT and T5, M5 and T5, and MT and TT (p < 0.0001).
Conclusion: Our data showed favorable primary implant stability (PS) values for both implant macro-geometries. Furthermore, the magnetodynamic preparation technique appears to be more effective in achieving higher PS values in low-density bone.
Keywords: implant macro-design; implant site preparation; insertion torque; magnetic mallet; magnetodynamic osteotomy; primary implant stability; resonance frequency analysis.
Conflict of interest statement
The authors declare no conflict of interest.
Figures







Similar articles
-
Primary Stability Optimization by Using Fixtures with Different Thread Depth According To Bone Density: A Clinical Prospective Study on Early Loaded Implants.Materials (Basel). 2019 Jul 27;12(15):2398. doi: 10.3390/ma12152398. Materials (Basel). 2019. PMID: 31357620 Free PMC article.
-
The effect of implant design and bone quality on insertion torque, resonance frequency analysis, and insertion energy during implant placement in low or low- to medium-density bone.Int J Prosthodont. 2015 Jan-Feb;28(1):40-7. doi: 10.11607/ijp.4063. Int J Prosthodont. 2015. PMID: 25588172
-
Enhancing primary implant stability by undersizing implant site preparation: A human cadaver study.J Stomatol Oral Maxillofac Surg. 2020 Feb;121(1):58-62. doi: 10.1016/j.jormas.2019.04.016. Epub 2019 May 2. J Stomatol Oral Maxillofac Surg. 2020. PMID: 31055093
-
Relative Contribution of Trabecular and Cortical Bone to Primary Implant Stability: An In vitro Model Study.J Pharm Bioallied Sci. 2024 Dec;16(Suppl 4):S3427-S3429. doi: 10.4103/jpbs.jpbs_899_24. Epub 2024 Oct 29. J Pharm Bioallied Sci. 2024. PMID: 39927043 Free PMC article.
-
The Influence of Implant Macro-geometry in Primary Stability in Low-Density Bone: An in vitro Study.J Int Soc Prev Community Dent. 2025 Apr 30;15(2):134-143. doi: 10.4103/jispcd.jispcd_155_24. eCollection 2025 Mar-Apr. J Int Soc Prev Community Dent. 2025. PMID: 40433430 Free PMC article.
Cited by
-
"Compression Necrosis" - A Cause of Concern for Early Implant Failure? Case Report and Review of Literature.Clin Cosmet Investig Dent. 2024 Mar 7;16:43-52. doi: 10.2147/CCIDE.S453798. eCollection 2024. Clin Cosmet Investig Dent. 2024. PMID: 38469322 Free PMC article.
-
Primary Stability of Implants Inserted into Polyurethane Blocks: Micro-CT and Analysis In Vitro.Bioengineering (Basel). 2024 Apr 15;11(4):383. doi: 10.3390/bioengineering11040383. Bioengineering (Basel). 2024. PMID: 38671804 Free PMC article.
-
Breakthroughs in Oral and Maxillofacial Surgery.J Clin Med. 2024 Jan 24;13(3):685. doi: 10.3390/jcm13030685. J Clin Med. 2024. PMID: 38337379 Free PMC article.
-
Using optimization approach to design dental implant in three types of bone quality - A finite element analysis.J Dent Sci. 2025 Jan;20(1):126-136. doi: 10.1016/j.jds.2024.09.017. Epub 2024 Oct 2. J Dent Sci. 2025. PMID: 39873028 Free PMC article.
-
Positional Accuracy of Dental Implants Placed by Means of Fully Guided Technique in Partially Edentulous Patients: A Retrospective Study.Clin Exp Dent Res. 2025 Jun;11(3):e70144. doi: 10.1002/cre2.70144. Clin Exp Dent Res. 2025. PMID: 40387371 Free PMC article.
References
LinkOut - more resources
Full Text Sources