. 2023 Sep 24;11(10):227.
doi: 10.3390/dj11100227.

Effect of Implant Macro-Design and Magnetodynamic Surgical Preparation on Primary Implant Stability: An In Vitro Investigation

Affiliations

Effect of Implant Macro-Design and Magnetodynamic Surgical Preparation on Primary Implant Stability: An In Vitro Investigation

Alessandro Antonelli et al. Dent J (Basel). .

Abstract

Background: Macro-geometry and surgical implant site preparation are two of the main factors influencing implant stability and potentially determining loading protocol. The purpose of this study was to assess the initial stability of various implant macro-designs using both magnetodynamic and traditional osteotomy techniques in low-density bone. The parameters examined included peak insertion torque (PIT), implant stability quotient (ISQ), and peak removal torque (PRT).

Methods: Four groups of 34 implants each were identified in accordance with the surgery and implant shape: T5 group (Five implant and osteotomy using drills); M5 group (Five implant and magnetodynamic osteotomy using Magnetic Mallet); TT group (TiSmart implant and osteotomy with drills); and MT group (TiSmart implant and magnetodynamic osteotomy). Every implant was placed into a low-density bone animal model and scanned using CBCT. The PIT and PRT were digitally measured in Newton-centimeters (Ncm) using a torque gauge device. The ISQ was analyzed by conducting resonance frequency analysis.

Results: The PIT values were 25.04 ± 4.4 Ncm for T5, 30.62 ± 3.81 Ncm for M5, 30 ± 3.74 Ncm for TT, and 32.05 ± 3.55 Ncm for MT. The average ISQ values were 68.11 ± 3.86 for T5, 71.41 ± 3.69 for M5, 70.88 ± 3.08 for TT, and 73 ± 3.5 for MT. The PRT values were 16.47 ± 4.56 Ncm for T5, 26.02 ± 4.03 Ncm for M5, 23.91 ± 3.28 Ncm for TT, and 26.93 ± 3.96 Ncm for MT. Based on our data analysis using a t-test with α = 0.05, significant differences in PIT were observed between TT and T5 (p < 0.0001), M5 and T5 (p < 0.0001), and MT and TT (p = 0.02). Significant differences in the ISQ were found between TT and T5 (p = 0.001), M5 and T5 (p < 0.001), and MT and TT (p = 0.01). The PRT also exhibited significant differences between TT and T5, M5 and T5, and MT and TT (p < 0.0001).

Conclusion: Our data showed favorable primary implant stability (PS) values for both implant macro-geometries. Furthermore, the magnetodynamic preparation technique appears to be more effective in achieving higher PS values in low-density bone.

Keywords: implant macro-design; implant site preparation; insertion torque; magnetic mallet; magnetodynamic osteotomy; primary implant stability; resonance frequency analysis.

PubMed Disclaimer

Conflict of interest statement

The authors declare no conflict of interest.

Figures

Figure 1
Figure 1
‘Five’ (a) and ‘TiSmart’ (b) (Leader Medica, Padua) implant macro-design.
Figure 2
Figure 2
Traditional implant site preparation (a). Magnetodynamic implant site preparation (b).
Figure 3
Figure 3
Dental implant placement and peak insertion torque recording.
Figure 4
Figure 4
Registration of the implant stability quotient.
Figure 5
Figure 5
The box plots display the differences in terms of primary stability (PS) between the T5 and TT groups.
Figure 6
Figure 6
The box plots illustrate the differences in terms of primary stability (PS) between the M5 and T5 groups.
Figure 7
Figure 7
The box plots highlight the differences in primary stability parameters between the MT and TT groups.

Similar articles

Cited by

References

    1. Duong H.Y., Roccuzzo A., Stähli A., Salvi G.E., Lang N.P., Sculean A. Oral health-related quality of life of patients rehabilitated with fixed and removable implant-supported dental prostheses. Periodontol. 2000. 2022;88:201–237. doi: 10.1111/prd.12419. - DOI - PMC - PubMed
    1. Pandey C., Rokaya D., Bhattarai B.P. Contemporary Concepts in Osseointegration of Dental Implants: A Review. Biomed. Res. Int. 2022;2022:6170452. doi: 10.1155/2022/6170452. - DOI - PMC - PubMed
    1. Sakka S., Baroudi K., Nassani M.Z. Factors associated with early and late failure of dental implants. J. Investig. Clin. Dent. 2012;3:258–261. doi: 10.1111/j.2041-1626.2012.00162.x. - DOI - PubMed
    1. Monje A., Ravidà A., Wang H.L., Helms J.A., Brunski J.B. Relationship Between Primary/Mechanical and Secondary/Biological Implant Stability. Int. J. Oral Maxillofac. Implant. 2019;34:s7–s23. doi: 10.11607/jomi.19suppl.g1. - DOI - PubMed
    1. Turkyilmaz I., Sennerby L., McGlumphy E.A., Tözüm T.F. Biomechanical aspects of primary implant stability: A human cadaver study. Clin. Implant Dent. Relat. Res. 2009;11:113–119. doi: 10.1111/j.1708-8208.2008.00097.x. - DOI - PubMed

LinkOut - more resources