Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Apr 16;2(1):161-173.
doi: 10.5194/mr-2-161-2021. eCollection 2021.

The decay of the refocused Hahn echo in double electron-electron resonance (DEER) experiments

Affiliations

The decay of the refocused Hahn echo in double electron-electron resonance (DEER) experiments

Thorsten Bahrenberg et al. Magn Reson (Gott). .

Abstract

Double electron-electron resonance (DEER) is a pulse electron paramagnetic resonance (EPR) technique that measures distances between paramagnetic centres. It utilizes a four-pulse sequence based on the refocused Hahn spin echo. The echo decays with increasing pulse sequence length 2(τ1+τ2), where τ1 and τ2 are the two time delays. In DEER, the value of τ2 is determined by the longest inter-spin distance that needs to be resolved, and τ1 is adjusted to maximize the echo amplitude and, thus, sensitivity. We show experimentally that, for typical spin centres (nitroxyl, trityl, and Gd(III)) diluted in frozen protonated solvents, the largest refocused echo amplitude for a given τ2 is obtained neither at very short τ1 (which minimizes the pulse sequence length) nor at τ1=τ2 (which maximizes dynamic decoupling for a given total sequence length) but rather at τ1 values smaller than τ2. Large-scale spin dynamics simulations based on the coupled cluster expansion (CCE), including the electron spin and several hundred neighbouring protons, reproduce the experimentally observed behaviour almost quantitatively. They show that electron spin dephasing is driven by solvent protons via the flip-flop coupling among themselves and their hyperfine couplings to the electron spin.

PubMed Disclaimer

Conflict of interest statement

The authors declare that they have no conflict of interest.

Figures

Figure 1
Figure 1
Schematics of (a) the three-pulse DEER sequence and (b) the dead-time-free four-pulse DEER sequence. Observer pulses are in blue and pump pulses in orange. The dashed line indicates t=0 .
Figure 2
Figure 2
Chemical structures of the paramagnetic centres studied in this work, namely (a) 3-maleimido-proxyl, (b) trityl OXO63, and (c) Gd-C2.
Figure 3
Figure 3
Refocused echo decay for 100  µM 3-maleimido-proxyl in H2O /  glycerol ( 80:20 v/v ) at 25 K. Panel (a) shows the echo amplitude as a function of τ1 and τ2 , and panel (b) shows the same data after normalization of each slice along τ1 . The red lines (a, b) indicate the location of the maxima along τ1 for fixed τ2 (upper line) and vice versa (lower line; only in a). Panel (c) shows slices along τ1 for several τ2 values (indicated by grey arrows in a, b), together with a Hahn echo decay.
Figure 4
Figure 4
Refocused echo decay for 100  µM 3-maleimido-proxyl in D 2 O  /  glycerol-d 8 ( 80:20 v/v ) at 25 K. Panel (a) shows the echo amplitude as a function of τ1 and τ2 , and panel (b) shows the same data after normalization of each slice along τ1 . Panel (c) shows slices along τ1 for several τ2 values (indicated by grey arrows in a, b), together with a Hahn echo decay.
Figure 5
Figure 5
Refocused echo decay for 100  µM 3-maleimido-proxyl in H2O /  glycerol solvents ( 80:20 v/v ) with varying degrees of solvent protonation at 25 K, i.e. (a) 25 %, (b) 50 %, and (c) 75 %. The red lines indicate the location of the maxima along τ1 for fixed τ2 (upper line) and vice versa (lower line). Plots for 100 % and 0 % solvent protonation are shown in Figs. 3a and 4a, respectively.
Figure 6
Figure 6
Refocused echo decays for 100  µM trityl OXO63 in (a)  H2O /  glycerol ( 80:20 v/v ) and (b)  D2O /  glycerol-d 8 at 25 K; in panel (a), the red lines show the position of the echo maximum along τ1 for each constant τ2 (upper line) and vice versa (lower line). Panel (c) shows slices of the data in panel (b) along τ1 for several τ2 values (indicated by grey arrows in b), together with a Hahn echo decay.
Figure 7
Figure 7
Refocused echo decays for 100  µM GdCl 3 in (a)  H2O /  glycerol ( 80:20 v/v ) and (b)  D2O /  glycerol-d 8 at 10 K; in panel (a), the red lines show the position of the echo maximum along τ1 for each constant τ2 (upper line) and vice versa (lower line). Grey arrows in (b) indicate the τ2 values of slices shown in (c), together with a Hahn echo decay.
Figure 8
Figure 8
Refocused echo decay experiments (constant τ2 ; variable τ1 ) on 25  µM MdfA V44C/V307C, doubly labelled with Gd-C2, at 10 K. Selected values for τ2 in nanoseconds are colour coded. Data taken from the Supplement of Yardeni et al. (2019).
Figure 9
Figure 9
Simulation of the refocused echo decay for 3-maleimido-proxyl in H2O /  glycerol ( 80:20 v/v ). Panel (a) shows the simulated refocused echo amplitude as a function of τ1 and τ2 , using 3-CCE. The τ1=τ2 line is shown in black. The upper red curve in panel (a) indicates the ridge of panel (b), which normalizes each slice along τ1 (with constant τ2 ) to unit maximal amplitude. The lower red curve in panel (a) is the analogous ridge for normalization along τ2 . Panel (c) shows the simulated refocused echo decay for τ1=τ2 as a function of cluster truncation level (1-CCE through 4-CCE).
Figure 10
Figure 10
Simulated refocused echo decay of 3-maleimido-proxyl. (a) Factorization of the four-cluster decay into a second-order term in bnm (top) and all higher-order terms (bottom). (b) Simulated refocused echo decays at two-, three-, and four-cluster level (2-CCE, 3-CCE, and 4-CCE; top), with the corresponding slice-wise normalized decays (bottom). For this simulation, a single orientation of the radical was used.

References

    1. Bahrenberg T, Yang Y, Goldfarb D, Feintuch A. rDEER: A Modified DEER Sequence for Distance Measurements Using Shaped Pulses. Magnetochemistry. 2019;5:20. doi: 10.3390/magnetochemistry5010020. - DOI
    1. Bahrenberg T, Jahn SM, Feintuch A, Stoll S, Goldfarb D. 2021. The decay of the refocused Hahn echo in DEER experiments, Raw data. Zenodo. - DOI - PMC - PubMed
    1. Borbat PP, Georgieva ER, Freed JH. Improved sensitivity for long-distance measurements in biomolecules: Five-pulse double electron-electron resonance. J Phys Chem Lett. 2013;4:170–175. doi: 10.1021/jz301788n. - DOI - PMC - PubMed
    1. Breitgoff FD, Soetbeer J, Doll A, Jeschke G, Polyhach YO. Artefact suppression in 5-pulse double electron electron resonance for distance distribution measurements. Phys Chem Chem Phys. 2017;19:15766–15779. doi: 10.1039/c7cp01488k. - DOI - PubMed
    1. Brown I. In: Time Domain Electron Spin Resonance. Kevan L, Schwartz RN, editors. John Wiley & Sons; New York: 1979. Electron spin-echo studies of relaxation processes in molecular solids; pp. 195–229.

LinkOut - more resources