. 2024;53(2):123-129.
doi: 10.1159/000534459. Epub 2023 Nov 8.

Removal of α1-Microglobulin Using Post-Dilution Online Hemodiafiltration with Polymethylmethacrylate Membrane: An Open-Label, Single-Arm Study

Affiliations

Removal of α1-Microglobulin Using Post-Dilution Online Hemodiafiltration with Polymethylmethacrylate Membrane: An Open-Label, Single-Arm Study

Shiori Yoshida et al. Blood Purif. 2024.

Abstract

Introduction: The removal of low- and medium-molecular-weight proteins has been improved with online hemodiafiltration (OL-HDF) and hemodialysis using high-flux membranes; however, the outcomes of patients with end-stage kidney disease (ESKD) undergoing dialysis treatment are still worse than in the general population. α1-Microglobulin (α1-m), with a molecular weight of 33,000 Da, may contribute to dialysis-related disorders and mortality. However, the removal is insufficient even with current OL-HDF using the polysulfone (PS) membrane, which is common in Japan. Polymethylmethacrylate (PMMA) membranes can remove medium- to high-molecular-weight proteins by adsorption. This study aimed to assess the efficacy of removing medium- to high-molecular-weight proteins, such as α1-m and β2-microglobulin (β2-m), through post-dilution OL-HDF with PMMA (Post-PMMA). The assessment was conducted in comparison to pre-dilution OL-HDF with PS (Pre-PS), using an open-label, single-arm study.

Methods: Seven patients with ESKD on Pre-PS underwent Post-PMMA with replacement volume of 30 mL/min (low flow) and 50 mL/min (high flow). Clearance and removal rates of α1-m, β2-m, small molecules, inflammatory cytokines, and albumin were measured at 60 and 240 min of treatment.

Results: Clearance rates of α1-m at 60 min were -2.8 ± 5.2 mL/min with Pre-PS, -0.4 ± 2.6 mL/min with Post-PMMA (low), and 0.6 ± 3.4 mL/min with Post-PMMA (high). The removal rate of α1-m was higher in Post-PMMA than that in Pre-HDF-PS (Post-PMMA [high] 17.7 ± 5.9%, Post-PMMA [low] 15.0 ± 5.6%, and Pre-PS 4.1 ± 5.5%). Adsorption clearance of β2-m was increased with Post-PMMA. Albumin leakage in Post-PMMA was not higher than that in Pre-PS.

Conclusion: The removal rate of α1-m with Post-PMMA was higher than that with Pre-PS. The PMMA membrane adsorbed β2-m, suggesting the removal effect of medium- to high-molecular-weight proteins by the adsorption method. Since Post-PMMA effectively removes α1-m without excessive albumin leakage, it will be useful for patients with ESKD, especially those with a poor nutritional status.

Keywords: Adsorption; Polymethylmethacrylate; Post-dilution online hemodiafiltration; α1-Microglobulin; β2-Microglobulin.

PubMed Disclaimer

Conflict of interest statement

This study was a commissioned study from Toray Medical, Inc. to Niigata University. The authors declare no other potential competing financial interests.