Observational Study
. 2023 Dec;116(12):580-589.
doi: 10.1016/j.acvd.2023.10.006. Epub 2023 Oct 26.

Safety and efficacy of the Amplatzer™ Trevisio™ intravascular delivery system: Post-approval study results

Affiliations
Free article
Observational Study

Safety and efficacy of the Amplatzer™ Trevisio™ intravascular delivery system: Post-approval study results

Sebastien Hascoet et al. Arch Cardiovasc Dis. 2023 Dec.
Free article

Abstract

Background: The Amplatzer™ Trevisio™ Intravascular Delivery System (Trevisio DS; Abbott Laboratories, Chicago, IL, USA) facilitates the delivery of Amplatzer™ Occluders and features an ultraflexible tip, which improves assessment of occluder position before release.

Aims: To assess the safety and efficacy of the Trevisio DS for transcatheter closure of patent foramen ovale and atrial septal defect.

Methods: The Amplatzer™ Trevisio™ Intravascular Delivery System Post-Approval Study was a prospective, postmarket, single-arm, multicentre, observational study of the Trevisio DS. Enrolled patients were indicated for transcatheter closure of patent foramen ovale or atrial septal defect. In all procedures, the Trevisio DS was used to deliver Amplatzer™ Occluders. Technical success was defined as successful deployment and release of at least one occluder. Device- or procedure-related serious adverse events were tracked until discharge or day 7, whichever occurred earlier.

Results: The study enrolled 144 patients with patent foramen ovale and 107 patients with atrial septal defect at 22 European sites; 53 patients with atrial septal defect (49.6%) were aged<18years. The rate of technical success was 98.4% (97.2% for atrial septal defect, 99.3% for patent foramen ovale). There was one serious adverse event (0.4%), an acute periprocedural device embolization that occurred after occluder release in a patient with atrial septal defect; the device was retrieved percutaneously. This was determined by the implanter to be unrelated to the performance of the Trevisio DS.

Conclusions: The Trevisio DS exhibited a high rate of technical success and an excellent safety profile during transcatheter closure of patent foramen ovale and atrial septal defect.

Keywords: Atrial septal defect; Clinical trial; Patent foramen ovale; Septal occluder device; Treatment outcome.