Volumetric effect of shunt adjustments in normal pressure hydrocephalus: a randomized, double-blind trial
- PMID: 37976516
- DOI: 10.3171/2023.9.JNS23668
Volumetric effect of shunt adjustments in normal pressure hydrocephalus: a randomized, double-blind trial
Abstract
Objective: MRI volumetry could be used as an alternative to invasive tests of shunt function. In this study, the authors aimed to assess the difference in ventricular volume (VV) before and after surgery and at different performance levels (PLs) of the shunt.
Methods: This study was a randomized, double-blind trial with a crossover design. The study sample consisted of 36 patients (25 men, 11 women) with a median age of 76 years. All patients had idiopathic normal pressure hydrocephalus (iNPH) and received a Strata shunt at the regional hospital in Östersund, Sweden, with an initial PL of 1.5. Participants underwent MRI with volumetric sequences before surgery and four times postoperatively: at 1 month before randomization to either PL 1.0 (n = 15) or 2.5 (n = 17); at 2 months before crossover to PL 2.5 or 1.0; at 3 months before lowering the PL to 0.5; and finally, at 3 months and 1 day after surgery before resetting the PL to 1.5. VV was measured semiautomatically using quantitative MRI. Both the patient and the examiner of clinical tests and volumetry were blinded to the PL.
Results: VV changed significantly between the presurgical level (median 129 ml) and the different shunt settings, i.e., PL 1.0 (median 115 ml), 1.5 (median 120 ml), and 2.5 (median 128 ml; p < 0.001). A unidirectional change in VV was observed for all participants between PL 1.0 and PL 2.5 (median 12 ml, range 2.1-40.7 ml, p < 0.001). No significant change was noted in VV after 24 hours at PL 0.5. Eight participants had asymptomatic subdural effusions at PL 1.0.
Conclusions: The consistent decrease in VV after shunt surgery and between PL 2.5 and 1.0 supports the idea that MRI volumetry could be a noninvasive method for evaluating shunt function in iNPH, preventing unnecessary shunt revisions. However, further studies on retest variability of VV as well as verification against advanced testing of shunt function are needed before a clinical implementation of this method can be performed. Clinical trial registration no.: NCT04599153 (ClinicalTrials.gov).
Keywords: cerebral ventricles; cerebrospinal fluid shunt; idiopathic normal pressure hydrocephalus; magnetic resonance imaging; performance level; shunt adjustment; ventricular volume.
Similar articles
-
Decrease of excessive daytime sleepiness after shunt treatment for normal pressure hydrocephalus.J Sleep Res. 2025 Apr;34(2):e14333. doi: 10.1111/jsr.14333. Epub 2024 Sep 14. J Sleep Res. 2025. PMID: 39275945 Free PMC article.
-
Endoscopic third ventriculostomy (ETV) for idiopathic normal pressure hydrocephalus (iNPH).Cochrane Database Syst Rev. 2015 Jul 29;2015(7):CD010033. doi: 10.1002/14651858.CD010033.pub2. Cochrane Database Syst Rev. 2015. PMID: 26222251 Free PMC article.
-
Artificial Intelligence for Automatic Analysis of Shunt Treatment in Presurgery and Postsurgery Computed Tomography Brain Scans of Patients With Idiopathic Normal Pressure Hydrocephalus.Neurosurgery. 2024 Dec 1;95(6):1329-1337. doi: 10.1227/neu.0000000000003015. Epub 2024 Jun 6. Neurosurgery. 2024. PMID: 38842320
-
Gait characteristics in idiopathic normal pressure hydrocephalus: a review on the effects of CSF tap test and shunt surgery.Eur J Med Res. 2024 Dec 30;29(1):633. doi: 10.1186/s40001-024-02162-2. Eur J Med Res. 2024. PMID: 39734225 Free PMC article.
-
Repeated lumbar or ventricular punctures in newborns with intraventricular haemorrhage.Cochrane Database Syst Rev. 2017 Apr 6;4(4):CD000216. doi: 10.1002/14651858.CD000216.pub2. Cochrane Database Syst Rev. 2017. PMID: 28384379 Free PMC article.
Cited by
-
Decrease of excessive daytime sleepiness after shunt treatment for normal pressure hydrocephalus.J Sleep Res. 2025 Apr;34(2):e14333. doi: 10.1111/jsr.14333. Epub 2024 Sep 14. J Sleep Res. 2025. PMID: 39275945 Free PMC article.
Publication types
MeSH terms
Associated data
LinkOut - more resources
Full Text Sources
Medical