. 2023 Dec;5(12):e905-e916.
doi: 10.1016/S2589-7500(23)00199-1.

A deep learning system for detection of early Barrett's neoplasia: a model development and validation study

Collaborators, Affiliations
Free article

A deep learning system for detection of early Barrett's neoplasia: a model development and validation study

K N Fockens et al. Lancet Digit Health. 2023 Dec.
Free article

Abstract

Background: Computer-aided detection (CADe) systems could assist endoscopists in detecting early neoplasia in Barrett's oesophagus, which could be difficult to detect in endoscopic images. The aim of this study was to develop, test, and benchmark a CADe system for early neoplasia in Barrett's oesophagus.

Methods: The CADe system was first pretrained with ImageNet followed by domain-specific pretraining with GastroNet. We trained the CADe system on a dataset of 14 046 images (2506 patients) of confirmed Barrett's oesophagus neoplasia and non-dysplastic Barrett's oesophagus from 15 centres. Neoplasia was delineated by 14 Barrett's oesophagus experts for all datasets. We tested the performance of the CADe system on two independent test sets. The all-comers test set comprised 327 (73 patients) non-dysplastic Barrett's oesophagus images, 82 (46 patients) neoplastic images, 180 (66 of the same patients) non-dysplastic Barrett's oesophagus videos, and 71 (45 of the same patients) neoplastic videos. The benchmarking test set comprised 100 (50 patients) neoplastic images, 300 (125 patients) non-dysplastic images, 47 (47 of the same patients) neoplastic videos, and 141 (82 of the same patients) non-dysplastic videos, and was enriched with subtle neoplasia cases. The benchmarking test set was evaluated by 112 endoscopists from six countries (first without CADe and, after 6 weeks, with CADe) and by 28 external international Barrett's oesophagus experts. The primary outcome was the sensitivity of Barrett's neoplasia detection by general endoscopists without CADe assistance versus with CADe assistance on the benchmarking test set. We compared sensitivity using a mixed-effects logistic regression model with conditional odds ratios (ORs; likelihood profile 95% CIs).

Findings: Sensitivity for neoplasia detection among endoscopists increased from 74% to 88% with CADe assistance (OR 2·04; 95% CI 1·73-2·42; p<0·0001 for images and from 67% to 79% [2·35; 1·90-2·94; p<0·0001] for video) without compromising specificity (from 89% to 90% [1·07; 0·96-1·19; p=0·20] for images and from 96% to 94% [0·94; 0·79-1·11; ] for video; p=0·46). In the all-comers test set, CADe detected neoplastic lesions in 95% (88-98) of images and 97% (90-99) of videos. In the benchmarking test set, the CADe system was superior to endoscopists in detecting neoplasia (90% vs 74% [OR 3·75; 95% CI 1·93-8·05; p=0·0002] for images and 91% vs 67% [11·68; 3·85-47·53; p<0·0001] for video) and non-inferior to Barrett's oesophagus experts (90% vs 87% [OR 1·74; 95% CI 0·83-3·65] for images and 91% vs 86% [2·94; 0·99-11·40] for video).

Interpretation: CADe outperformed endoscopists in detecting Barrett's oesophagus neoplasia and, when used as an assistive tool, it improved their detection rate. CADe detected virtually all neoplasia in a test set of consecutive cases.

Funding: Olympus.

PubMed Disclaimer

Conflict of interest statement

Declaration of interests MB is a consultant for Medtronic and a board member for Norgine and Ambu, and reports funding from Pentax Medical and medical training from Olympus. JJB reports financial support for institutional review board-approved research from C2Therapeutics and Pentax Medical, Medtronic, and Aqua Medical. TB reports consulting fees from Olympus, Boston Scientific, and Microtech; lecture fees from Olympus, Fujifilm, Pentax, Microtech, ERBE, and Medtronic; payment for expert testimony from Olympus; and participation on a data safety monitoring board or advisory board at Olympus. RCM-H reports consultant fees from Janssen. OP reports speaker honorarium from Boston Scientific, Medtronic, Fujifilm, Olympus, Aohua, Falk, and BMS. REP reports consulting fees from Medtronic and Microtech and speaker fees from Pentax. KR reports consulting fees from Olympus and lecture fees from Olympus. BLAMW reports financial support from Pentax Medical and St Antonius Research Fund; research support from Aqua Medical; consulting fees from Pentax Medical; and speaker fees from Pentax Medical and is chair of the European Society of Gastrointestinal Endoscopy guideline working group for the revision of the Barrett's oesophagus guideline. All other authors declare no competing interests.