Black-Blood Magnetization Prepared 2 Rapid Acquisition Gradient Echoes: A Fast and Three-Dimensional MR Black-Blood T1 Mapping Technique for Quantitative Assessment of Atherosclerosis and Venous Thrombosis
- PMID: 38009385
- DOI: 10.1002/jmri.29156
Black-Blood Magnetization Prepared 2 Rapid Acquisition Gradient Echoes: A Fast and Three-Dimensional MR Black-Blood T1 Mapping Technique for Quantitative Assessment of Atherosclerosis and Venous Thrombosis
Abstract
Background: Blood flow signals may be a confounder in quantifying T1 values of plaque or thrombus and how to realize black-blood T1 mapping remains a challenge task.
Purpose: To develop a fast and three-dimensional black-blood T1 mapping technique for quantitative assessment of atherosclerosis and venous thrombosis.
Study type: Sequence development and optimization via phantoms and volunteers as well as pilot prospective.
Phantom and subjects: Numerical simulations, a standard phantom, 8 healthy volunteers (mean age, 22 ± 1 years; 5 males), and 19 patients (mean age, 57 ± 14 years; 13 males) with atherosclerosis or venous thrombosis.
Field strength/sequence: 3T/inversion recovery spin-echo sequence (IR-SE), magnetization prepared 2 rapid acquisition gradient echoes (MP2RAGE), and black-blood prepared MP2RAGE (BB-MP2RAGE).
Assessment: The black-blood preparation (i.e., delay alternating with nutation for tailored excitation, DANTE) was incorporated into MP2RAGE for black-blood T1 mapping. The BB-MP2RAGE was optimized numerically based on the Bloch equation, and then the phantom study was performed to verify the accuracy of T1 mapping by BB-MP2RAGE against IR-SE and MP2RAGE. Preliminary clinical validation was prospectively performed to assess the flow suppression effect and its potential application in plaque and thrombosis identification.
Statistical tests: Pearson correlation test, Bland-Altman analysis, paired t-test, and intraclass correlation coefficient. A P value <0.05 indicates a statistically significant difference.
Results: Phantom experiments showed comparable accuracy of T1 maps by BB-MP2RAGE with IR-SE and MP2RAGE (all r2 > 0.99); Compared to MP2RAGE, BB-MP2RAGE effectively nulled the blood flow signals, and had a significant improvement in contrast-to-noise ratio between static tissue and blood (250.5 ± 66.6 vs. 91.9 ± 35.9). BB-MP2RAGE can quantify plaque or thrombus T1 relaxation time with blood flow signal suppression.
Data conclusion: Accurate T1 mapping with sufficient blood flow suppression was achieved by BB-MP2RAGE. BB-MP2RAGE has the potential to quantitatively characterize atherosclerosis and venous thrombosis.
Level of evidence: 1 TECHNICAL EFFICACY: Stage 1.
Keywords: T1 mapping; atherosclerosis; black‐blood imaging; venous thrombosis.
© 2023 International Society for Magnetic Resonance in Medicine.
Similar articles
-
Optimization and validation of low-field MP2RAGE T1 mapping on 0.35T MR-Linac: Toward adaptive dose painting with hypoxia biomarkers.Med Phys. 2024 Nov;51(11):8124-8140. doi: 10.1002/mp.17353. Epub 2024 Aug 14. Med Phys. 2024. PMID: 39140821
-
Magnetization-prepared 2 Rapid Gradient-Echo MRI for B1 Insensitive 3D T1 Mapping of Hip Cartilage: An Experimental and Clinical Validation.Radiology. 2021 Apr;299(1):150-158. doi: 10.1148/radiol.2021200085. Epub 2021 Feb 23. Radiology. 2021. PMID: 33620288
-
Black-blood multicontrast imaging of carotid arteries with DANTE-prepared 2D and 3D MR imaging.Radiology. 2014 Nov;273(2):560-9. doi: 10.1148/radiol.14131717. Epub 2014 Jun 11. Radiology. 2014. PMID: 24918958
-
Black-blood thrombus imaging (BTI): a contrast-free cardiovascular magnetic resonance approach for the diagnosis of non-acute deep vein thrombosis.J Cardiovasc Magn Reson. 2017 Jan 18;19(1):4. doi: 10.1186/s12968-016-0320-8. J Cardiovasc Magn Reson. 2017. PMID: 28095878 Free PMC article.
-
Fast magnetization transfer saturation imaging of the brain using MP2RAGE T1 mapping.Magn Reson Med. 2024 Oct;92(4):1540-1555. doi: 10.1002/mrm.30143. Epub 2024 May 4. Magn Reson Med. 2024. PMID: 38703017
Cited by
-
A signal model for fat-suppressed T1-mapping and dynamic contrast-enhanced MRI with interrupted spoiled gradient-echo readout.NMR Biomed. 2025 Jan;38(1):e5289. doi: 10.1002/nbm.5289. Epub 2024 Nov 21. NMR Biomed. 2025. PMID: 39571186 Free PMC article.
References
-
- Qiao H, Li D, Cao J, et al. Quantitative evaluation of carotid atherosclerotic vulnerable plaques using in vivo T1 mapping cardiovascular magnetic resonaonce: Validation by histology. J Cardiovasc Magn Reson 2020;22(1):38. https://doi.org/10.1186/s12968-020-00624-0.
-
- Jiang Y, Zhu C, Peng W, et al. Ex‐vivo imaging and plaque type classification of intracranial atherosclerotic plaque using high resolution MRI. Atherosclerosis 2016;249:10‐16. https://doi.org/10.1016/j.atherosclerosis.2016.03.033.
-
- Saha P, Andia ME, Modarai B, et al. Magnetic resonance T1 relaxation time of venous thrombus is determined by iron processing and predicts susceptibility to lysis. Circulation 2013;128:729‐736. https://doi.org/10.1161/CIRCULATIONAHA.113.001371.
-
- Ahmadi A, Leipsic J, Blankstein R, et al. Do plaques rapidly Progress prior to myocardial infarction? The interplay between plaque vulnerability and progression. Circ Res 2015;117(1):99‐104. https://doi.org/10.1161/CIRCRESAHA.117.305637.
-
- Qi H, Sun J, Qiao H, et al. Carotid intraplaque hemorrhage imaging with quantitative Vessel Wall T1 mapping: Technical development and initial experience. Radiology 2018;287(1):276‐284. https://doi.org/10.1148/radiol.2017170526.
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical
Research Materials