Case Reports
. 2023 Jul 24;9(10):728-731.
doi: 10.1016/j.hrcr.2023.07.015. eCollection 2023 Oct.

T-wave oversensing precipitating cardiogenic shock in a pacer-dependent patient

Affiliations
Case Reports

T-wave oversensing precipitating cardiogenic shock in a pacer-dependent patient

Shubash Adhikari et al. HeartRhythm Case Rep. .
No abstract available

Keywords: Bradycardia; Cardiogenic shock; Inappropriate ventricular pacing; Pacemaker; T-wave oversensing.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Twelve-lead electrocardiogram showing atrial fibrillation and ventricular pacing at a rate of 48 beats/min.
Figure 2
Figure 2
Permanent pacemaker interrogation showing inappropriate ventricular sensing at the T waves, confirming T-wave oversensing that caused a delay in ventricular pacing.
Figure 3
Figure 3
Hemodynamics review chart showing the improvement of cardiac index (C.I.) from 1.68 to 2.36 L/min/m2 after abolition of the T-wave oversensing and restoration of an appropriate ventricular pacing.

References

    1. Francis J. T wave oversensing. Indian Pacing Electrophysiol J. 2010;10:236–238. - PMC - PubMed
    1. Bortnik M., Fundaliotis A., Schaffer A., Occhetta E., Marino P. Transient hyperkalemia-induced T-wave oversensing as a cause of inappropriate ICD discharges. J Cardiol Cases. 2013;8:164–167. - PMC - PubMed
    1. Chhabra L., Khalid N., Kluger J. Myocardial ischemic hyperacute T-wave oversensing leading to a defibrillator shock storm. Proc (Bayl Univ Med Cent) 2015;28:200–203. - PMC - PubMed
    1. Srivathsan K., Scott L.R., Altemose G.T. T-wave oversensing and inappropriate shocks: a case report. EP Europace. 2008;10:552–555. - PubMed
    1. Salgado B.C., Coram R., Mandrola J., Gopinathannair R. Inappropriate shock from delayed T-wave oversensing by a subcutaneous implantable cardioverter-defibrillator after septal myectomy for hypertrophic cardiomyopathy. HeartRhythm Case Rep. 2018;4:408–411. - PMC - PubMed

Publication types

LinkOut - more resources