Strategies to Minimize Access Site-related Complications in Patients Undergoing Transfemoral Artery Procedures with Large-bore Devices
- PMID: 38073100
- DOI: 10.2174/0115701611233184231206100222
Strategies to Minimize Access Site-related Complications in Patients Undergoing Transfemoral Artery Procedures with Large-bore Devices
Abstract
Large bore accesses refer to accesses with a diameter of 10 French or greater and are necessary for various medical devices, including those used in transcatheter aortic valve replacement, endovascular aneurysm repair stent-grafts, and percutaneous mechanical support devices. Notably, the utilization of these devices via femoral access is steadily increasing due to advancements in technology and implantation techniques, which are expanding the pool of patients suitable for percutaneous procedures. However, procedures involving large bore devices carry a high risk of bleeding and vascular complications (VCs), impacting both morbidity and long-term mortality. In this review article, we will first discuss the incidence, determinants, and prognostic impact of VCs in patients undergoing large bore access procedures. Subsequently, we will explore the strategies developed in recent years to minimize VCs, including techniques for optimizing vascular puncture through femoral cannulation, such as the use of echo-guided access cannulation and fluoroscopic guidance. Additionally, we will evaluate existing vascular closure devices designed for large bore devices. Finally, we will consider new pharmacological strategies aimed at reducing the risk of periprocedural access-related bleeding.
Keywords: Large bore access procedures; artery procedures; femoral access; long-term mortality.; morbidity; vascular complications.
Copyright© Bentham Science Publishers; For any queries, please email at epub@benthamscience.net.
References
-
- Sorrentino S.; Giustino G.; Moalem K.; Indolfi C.; Mehran R.; Dangas G.D.; Antithrombotic treatment after transcatheter heart valves implant. Semin Thromb Hemost 2018,44(1),38-45 - DOI
-
- Cheney A.E.; McCabe J.M.; Alternative percutaneous access for large bore devices. Circ Cardiovasc Interv 2019,12(6),e007707 - DOI
Publication types
MeSH terms
LinkOut - more resources